Kilovoltage beam model for flat panel imaging system with bow-tie filter for scatter prediction and correction.

M Blessing1, M S Bhagwat, Y Lyatskaya, J R Bellon, J Hesser, P Zygmanski.   

Abstract

PURPOSE: Kilovoltage flat-panel imaging systems are used for cone-beam Computed Tomography (CBCT) and digital Tomosynthesis (DTS). Hereby, the presence of scatter and relatively large dose from imaging are challenging factors. In this study a phenomenological beam model was developed to characterize imager response to imaging beams with a bow-tie filter (Varian OBI system). MATERIALS AND
METHOD: The kilovoltage beam model was based on dose ratio formalism and thus was using standard concepts of megavoltage dose calculation such as scatter factors, tissue maximum ratio and off-axis ratio. Primary and scatter (head and phantom scatter) were modeled with three Gaussian kernels. Parameters were based on measured transmission images for slabs of solid water of different total thickness and various jaw settings.
RESULTS: The beam model was used to evaluate contributions from primary, secondary and tertiary contributions for different geometrical objects such as cylinders and step-like phantoms. Theoretical predictions of radiographs using the model for known objects are consistent with the measurements.
CONCLUSION: Secondary and tertiary contributions were interpreted as scatter and can be subtracted from CBCT projections based on the analytical model. Therefore our model can provide a basis for improvement of image quality (less artifacts due to scatter, better contrast and resolution) in CBCT reconstruction. Copyright Â
© 2011 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21570884     DOI: 10.1016/j.ejmp.2011.04.001

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  2 in total

1.  Narrow-Energy-Width CT Based on Multivoltage X-Ray Image Decomposition.

Authors:  Jiaotong Wei; Yan Han; Ping Chen
Journal:  Int J Biomed Imaging       Date:  2017-11-07

2.  Dynamic bowtie filter for cone-beam/multi-slice CT.

Authors:  Fenglin Liu; Qingsong Yang; Wenxiang Cong; Ge Wang
Journal:  PLoS One       Date:  2014-07-22       Impact factor: 3.240

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.