Isotope dilution-LC-MS/MS reference method assessment of serum folate assay accuracy and proficiency testing consensus mean.

Sheena Blackmore1, Christine M Pfeiffer, Anne Lee, Zia Fazili, Malcolm S Hamilton.   

Abstract

BACKGROUND: Current methods for measuring folates in clinical laboratories are competitive folate binding protein assays. These assays show a considerable lack of agreement that has implications for the comparability of data across studies as well as for long-term population studies. The development of isotope dilution-liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS) reference methods permitted the evaluation of method accuracy and consistency over time.
METHODS: We measured 3 pools of human serum by ID-LC-MS/MS, calculated values for total folate, and distributed the same pools to participants in a national External Quality Assessment scheme. We used linear regression to compare the all-laboratory and method data with reference method values. The exercise was repeated after 18 months to assess the stability of the all-laboratory and method data.
RESULTS: The distributed serum pools had mass spectrometry values for folate species typical of those found in healthy individuals from populations not receiving dietary folic acid fortification. There was good agreement of the all-laboratory data set with the reference method (y =0.86x + 0.91 μg/L) at both time points. Linear regression demonstrated that the Abbott Architect showed the closest agreement with the reference method. The Roche Elecsys method was nonlinear and showed a calibration offset of 2.6 μg/L (4.57 nmol/L).
CONCLUSIONS: Calibration of serum folate assays traceable to higher-order reference methods increases method accuracy and improves consistency. The all-laboratory consensus mean proved sufficiently accurate and stable to be used as the target for monitoring laboratory performance.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21561849     DOI: 10.1373/clinchem.2010.160135

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  3 in total

1.  Changes in measurement procedure from a radioassay to a microbiologic assay necessitate adjustment of serum and RBC folate concentrations in the U.S. population from the NHANES 1988-2010.

Authors:  Christine M Pfeiffer; Jeffery P Hughes; Ramon A Durazo-Arvizu; David A Lacher; Christopher T Sempos; Mindy Zhang; Elizabeth A Yetley; Clifford L Johnson
Journal:  J Nutr       Date:  2012-03-21       Impact factor: 4.798

Review 2.  Framework for laboratory harmonization of folate measurements in low- and middle-income countries and regions.

Authors:  Christine M Pfeiffer; Mindy Zhang; Shameem Jabbar
Journal:  Ann N Y Acad Sci       Date:  2018-01-29       Impact factor: 5.691

Review 3.  Biomarkers of Nutrition for Development-Folate Review.

Authors:  Lynn B Bailey; Patrick J Stover; Helene McNulty; Michael F Fenech; Jesse F Gregory; James L Mills; Christine M Pfeiffer; Zia Fazili; Mindy Zhang; Per M Ueland; Anne M Molloy; Marie A Caudill; Barry Shane; Robert J Berry; Regan L Bailey; Dorothy B Hausman; Ramkripa Raghavan; Daniel J Raiten
Journal:  J Nutr       Date:  2015-06-03       Impact factor: 4.798

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.