Biomedical Ti-Mo alloys with surface machined and modified by laser beam: biomechanical, histological, and histometric analysis in rabbits.

Nilson T C Oliveira1, Fernando P S Guastaldi, Vittoria Perrotti, Eduardo Hochuli-Vieira, Antonio C Guastaldi, Adriano Piattelli, Giovanna Iezzi.   

Abstract

PURPOSE: In vivo bone response was assessed by removal torque, hystological and histometrical analysis on a recently developed biomedical Ti-15Mo alloy, after surface modification by laser beam irradiation, installed in the tibia of rabbits.
MATERIALS AND METHODS: A total of 32 wide cylindrical Ti-15Mo dental implants were obtained (10mm × 3.75mm). The implants were divided into two groups: 1) control samples (Machined surface - MS) and 2) implants with their surface modified by Laser beam-irradiation (Test samples - LS). Six implants of each surface were used for removal torque test and 10 of each surface for histological and histometrical analysis. The implants were placed in the tibial metaphyses of rabbits.
RESULTS: Average removal torque was 51.5Ncm to MS and >90Ncm to LS. Bone-to-implant-contact percentage was significantly higher for LS implants both in the cortical and marrow regions.
CONCLUSIONS: The present study demonstrated that laser treated Ti-15Mo alloys are promising materials for biomedical application.
© 2011 Wiley Periodicals, Inc.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21554530     DOI: 10.1111/j.1708-8208.2011.00354.x

Source DB:  PubMed          Journal:  Clin Implant Dent Relat Res        ISSN: 1523-0899            Impact factor:   3.932


  4 in total

1.  Sol-gel based calcium phosphates coatings deposited on binary Ti-Mo alloys modified by laser beam irradiation for biomaterial/clinical applications.

Authors:  Marcio Luiz Dos Santos; Carla Dos Santos Riccardi; Edson de Almeida Filho; Antonio C Guastaldi
Journal:  J Mater Sci Mater Med       Date:  2018-06-11       Impact factor: 3.896

2.  Ti-15Mo Alloy Decreases the Stress Concentration in Mandibular Angle Fracture Internal Fixation Hardware.

Authors:  F P S Guastaldi; A P Martini; E P Rocha; E Hochuli-Vieira; A C Guastaldi
Journal:  J Maxillofac Oral Surg       Date:  2019-06-03

3.  Behavior of rat bone marrow stem cells on titanium surfaces modified by laser-beam and deposition of calcium phosphate.

Authors:  F Florian; F P S Guastaldi; M A Cominotte; L C Pires; A C Guastaldi; J A Cirelli
Journal:  J Mater Sci Mater Med       Date:  2021-05-17       Impact factor: 3.896

4.  Plasma Electrolytic Oxidation as a Feasible Surface Treatment for Biomedical Applications: an in vivo study.