Comparison of software versions for body composition analysis using the PIXImus dual-energy X-ray absorptiometer.

M S Johnson1, N M Landy, E P Potter, T R Nagy.   

Abstract

We have previously validated the use of dual-energy X-ray absorptiometry (DXA) for measuring body composition of mice using the GE-Lunar PIXImus and software version 1.42 [1]. Since that report, newer versions of the software have been released. The purpose of the present study was to compare results from our original study with results analyzed using two newer versions of software (versions 1.44 and 1.45). Body composition data (lean tissue mass [LTM], fat mass [FM], bone mineral content [BMC], and bone mineral density [BMD]) were obtained from DXA scans of twenty-five, anesthetized male C57Bl/6J mice (6-11 weeks old; 19 to 29g). Relative to version 1.42, versions 1.44 and 1.45 significantly (P<0.001) overestimated LTM and BMD and underestimated FM and BMC. However, compared to carcass analysis, versions 1.44 and 1.45 significantly overestimated both FM and LTM and underestimated BMC. Results from 1.44 and 1.45 were highly correlated with carcass values for all body composition parameters. Prediction equations were developed for the two new software versions. Applying the prediction equation from 1.42, to the data obtained from 1.44 and 1.45 resulted in FM and LTM that were worse than if no equation was used. However, using their own developed equations resulted in data that were not significantly different than that from carcass analysis. These data suggest that software-specific equations are necessary for comparing DXA-derived data to that of chemical analysis.

Entities:  

Year:  2005        PMID: 21552433      PMCID: PMC3088425     

Source DB:  PubMed          Journal:  Int J Body Compos Res        ISSN: 1479-456X


  4 in total

1.  Precision and accuracy of dual-energy X-ray absorptiometry for determining in vivo body composition of mice.

Authors:  T R Nagy; A L Clair
Journal:  Obes Res       Date:  2000-08

2.  Validation and calibration of DEXA body composition in mice.

Authors:  Robert Brommage
Journal:  Am J Physiol Endocrinol Metab       Date:  2003-05-20       Impact factor: 4.310

3.  Bone loss detection in rats using a mouse densitometer.

Authors:  N Binkley; D B Dahl; J Engelke; T Kawahara-Baccus; D Krueger; R J Colman
Journal:  J Bone Miner Res       Date:  2003-02       Impact factor: 6.741

4.  Bone mineral measurements in mice: comparison of two devices.

Authors:  Sami Kolta; Marie-Christine De Vernejoul; Pierre Meneton; Jacques Fechtenbaum; Christian Roux
Journal:  J Clin Densitom       Date:  2003       Impact factor: 2.963

  4 in total
  2 in total

1.  Validation of quantitative magnetic resonance for the determination of body composition of mice.

Authors:  A S Jones; M S Johnson; T R Nagy
Journal:  Int J Body Compos Res       Date:  2009

2.  Premenopausal Singaporean Women Suffering from Major Depressive Disorder Treated with Selective Serotonin Reuptake Inhibitors Had Similar Bone Mineral Density as Compared with Healthy Controls.

Authors:  Roger C Ho; Anna N Chua; Syeda Fabeha Husain; Wanqiu Tan; Fengyi Hao; Giang T Vu; Bach X Tran; Hien Thu Nguyen; Roger S McIntyre; Cyrus S Ho
Journal:  Diagnostics (Basel)       Date:  2022-01-03
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.