Danko Tomasic, Bozidar Ferek-Petric, Sandro Brusich, Agostino P Accardo.
Abstract
Cardiac contractions bend the implanted cardiac lead body, extend and compress the lead conductors, their insulation and the inserted stylet. Magnitude of lead deflection depends on cardiac muscle contraction forces. The purpose of study was to measure the charge generated due to triboelectric effect between one of the lead conductors and the inserted stylet. The charge was measured by differential charge amplifier being connected to isolation amplifier and power supply. Sensor signal, ECG and intracardiac electrograms were acquired. Three models of custom designed leads were implanted in 8 sheep. Measurements were done in 18 patients undergoing pacemaker implantation and replacement procedures. Atrial and ventricular tensiometric signals were recorded in dual chamber and in single-lead VDD patients. Recordings in sinus rhythm at various AV intervals and in supraventricular tachycardia were done. In average, charge variation between 1 and 600 pC was measured. Tensiometric stylet could be feasible hemodynamic sensor for myocardial contraction detection. Its main advantage is that it is easily exchangeable and universal for all leads.Entities:
Mesh:
Year: 2011 PMID: 21536515 DOI: 10.1109/TBME.2011.2148198
Source DB: PubMed Journal: IEEE Trans Biomed Eng ISSN: 0018-9294 Impact factor: 4.538