Abstract
This rapid fluorescence polarization immunoassay for urinary vanillylmandelic acid (VMA) involves use of our previously described antiserum and label and the program for 5-hydroxyindoleacetic acid in the Abbott TDx fluorimeter. Urine samples were measured directly, without pretreatment. The minimum detectable concentration was 0.3 mg/L, and the range of the standard curve was 0.3-200.0 mg/L. Precision, analytical recovery, and correlation of results with those by the Pisano method (Clin Chim Acta 1962;7:285-91) were all satisfactory. With this procedure one can determine the VMA concentration in 10 urine sample in 22 min. This is the first report of a clinical immunoassay for VMA and should greatly simplify screening for neural crest tumors.Entities:
Mesh:
Substances:
Year: 1990 PMID: 2153482
Source DB: PubMed Journal: Clin Chem ISSN: 0009-9147 Impact factor: 8.327