Literature DB >> 21529004

A novel algorithm for quick and continuous tracing the change of radon concentration in environment.

Yanliang Tan1, Detao Xiao.   

Abstract

Several measurements of the radon concentration are performed by RAD7 in the University of South China. We find that 30-40 min is needed for RAD7 for tracing the concentration of the standard radon chamber. There are two reasons. The first is that the sufficient time of air cycle is needed for the radon concentration in internal cell of RAD7 equal to that of the environment; and the second is that the sufficient decay time is needed for the (218)Po concentration in internal cell of RAD7 equal to that of the radon. We used a zeroth order approximation to describe the evolution of the environment radon concentration, and obtained a novel algorithm for quick and continuous tracing the change of radon concentration. The corrected radon concentration obtained through this method is in good agreement with the reference value. This method can be applied to develop and improve the instruments for tracing the change of radon concentration quickly.
© 2011 American Institute of Physics

Year:  2011        PMID: 21529004     DOI: 10.1063/1.3572271

Source DB:  PubMed          Journal:  Rev Sci Instrum        ISSN: 0034-6748            Impact factor:   1.523


  1 in total

1.  Measuring radon exhalation rate in two cycles avoiding the effects of back-diffusion and chamber leakage.

Authors:  Yanliang Tan; Detao Xiao
Journal:  Environ Monit Assess       Date:  2013-04-25       Impact factor: 2.513

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.