Dosimetry protocol for the forthcoming clinical trials in synchrotron stereotactic radiation therapy (SSRT).

Y Prezado1, M Vautrin, I Martínez-Rovira, A Bravin, F Estève, H Elleaume, P Berkvens, J F Adam.   

Abstract

PURPOSE: An adequate dosimetry protocol for synchrotron radiation and the specific features of the ID17 Biomedical Beamline at the European Synchrotron Radiation Facility are essential for the preparation of the forthcoming clinical trials in the synchrotron stereotactic radiation therapy (SSRT). The main aim of this work is the definition of a suitable protocol based on standards of dose absorbed to water. It must allow measuring the absolute dose with an uncertainty within the recommended limits for patient treatment of 2%-5%.
METHODS: Absolute dosimetry is performed with a thimble ionization chamber (PTW semiflex 31002) whose center is positioned at 2 g cm(-2) equivalent depth in water. Since the available synchrotron beam at the ESRF Biomedical Beamline has a maximum height of 3 mm, a scanning method was employed to mimic a uniform exposition of the ionization chamber. The scanning method has been shown to be equivalent to a broad beam irradiation. Different correction factors have been assessed by using Monte Carlo simulations.
RESULTS: The absolute dose absorbed to water at 80 keV was measured in reference conditions with a 2% global uncertainty, within the recommended limits. The dose rate was determined to be in the range between 14 and 18 Gy/min, that is to say, a factor two to three times higher than the 6 Gy/min achievable in RapidArc or VMAT machines. The dose absorbed to water was also measured in a RW3 solid water phantom. This phantom is suitable for quality assurance purposes since less than 2% average difference with respect to the water phantom measurements was found. In addition, output factors were assessed for different field sizes.
CONCLUSIONS: A dosimetry protocol adequate for the specific features of the SSRT technique has been developed. This protocol allows measuring the absolute dose absorbed to water with an accuracy of 2%. It is therefore satisfactory for patient treatment.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21520884     DOI: 10.1118/1.3556561

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  3 in total

1.  Radiochromic film dosimetry in synchrotron radiation breast computed tomography: a phantom study.

Authors:  Giovanni Mettivier; Marica Masi; Fulvia Arfelli; Luca Brombal; Pasquale Delogu; Francesca Di Lillo; Sandro Donato; Christian Fedon; Bruno Golosio; Piernicola Oliva; Luigi Rigon; Antonio Sarno; Angelo Taibi; Paolo Russo
Journal:  J Synchrotron Radiat       Date:  2020-04-22       Impact factor: 2.616

2.  Brain virtual histology with X-ray phase-contrast tomography Part I: whole-brain myelin mapping in white-matter injury models.

Authors:  Matthieu Chourrout; Hugo Rositi; Elodie Ong; Violaine Hubert; Alexandre Paccalet; Louis Foucault; Awen Autret; Barbara Fayard; Cécile Olivier; Radu Bolbos; Françoise Peyrin; Claire Crola-da-Silva; David Meyronet; Olivier Raineteau; Héléne Elleaume; Emmanuel Brun; Fabien Chauveau; Marlene Wiart
Journal:  Biomed Opt Express       Date:  2022-02-23       Impact factor: 3.732

3.  Evaluation of silicon strip detectors in transmission mode for online beam monitoring in microbeam radiation therapy at the Australian Synchrotron.

Authors:  Jeremy Davis; Andrew Dipuglia; Matthew Cameron; Jason Paino; Ashley Cullen; Susanna Guatelli; Marco Petasecca; Anatoly Rosenfeld; Michael Lerch
Journal:  J Synchrotron Radiat       Date:  2022-01-01       Impact factor: 2.616

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.