Evaluation of a CdTe semiconductor based compact γ camera for sentinel lymph node imaging.

Paolo Russo1, Assunta S Curion, Giovanni Mettivier, Michela Esposito, Michela Aurilio, Corradina Caracò, Luigi Aloj, Secondo Lastoria.   

Abstract

PURPOSE: The authors assembled a prototype compact gamma-ray imaging probe (MediPROBE) for sentinel lymph node (SLN) localization. This probe is based on a semiconductor pixel detector. Its basic performance was assessed in the laboratory and clinically in comparison with a conventional gamma camera.
METHODS: The room-temperature CdTe pixel detector (1 mm thick) has 256 x 256 square pixels arranged with a 55 microm pitch (sensitive area 14.08 x 14.08 mm2), coupled pixel-by-pixel via bump-bonding to the Medipix2 photon-counting readout CMOS integrated circuit. The imaging probe is equipped with a set of three interchangeable knife-edge pinhole collimators (0.94, 1.2, or 2.1 mm effective diameter at 140 keV) and its focal distance can be regulated in order to set a given field of view (FOV). A typical FOV of 70 mm at 50 mm skin-to-collimator distance corresponds to a minification factor 1:5. The detector is operated at a single low-energy threshold of about 20 keV.
RESULTS: For 99 mTc, at 50 mm distance, a background-subtracted sensitivity of 6.5 x 10(-3) cps/kBq and a system spatial resolution of 5.5 mm FWHM were obtained for the 0.94 mm pinhole; corresponding values for the 2.1 mm pinhole were 3.3 x 10(-2) cps/kBq and 12.6 mm. The dark count rate was 0.71 cps. Clinical images in three patients with melanoma indicate detection of the SLNs with acquisition times between 60 and 410 s with an injected activity of 26 MBq 99 mTc and prior localization with standard gamma camera lymphoscintigraphy.
CONCLUSIONS: The laboratory performance of this imaging probe is limited by the pinhole collimator performance and the necessity of working in minification due to the limited detector size. However, in clinical operative conditions, the CdTe imaging probe was effective in detecting SLNs with adequate resolution and an acceptable sensitivity. Sensitivity is expected to improve with the future availability of a larger CdTe detector permitting operation at shorter distances from the patient skin.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21520866     DOI: 10.1118/1.3555034

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  3 in total

1.  A Multimodality Hybrid Gamma-Optical Camera for Intraoperative Imaging.

Authors:  John E Lees; Sarah L Bugby; Mohammed S Alqahtani; Layal K Jambi; Numan S Dawood; William R McKnight; Aik H Ng; Alan C Perkins
Journal:  Sensors (Basel)       Date:  2017-03-09       Impact factor: 3.576

2.  Intraoperative biophotonic imaging systems for image-guided interventions.

Authors:  Salar Sajedi; Hamid Sabet; Hak Soo Choi
Journal:  Nanophotonics       Date:  2018-12-14       Impact factor: 8.449

Review 3.  Above and Beyond Robotic Surgery and 3D Modelling in Paediatric Cancer Surgery.

Authors:  Laura Privitera; Irene Paraboschi; Kate Cross; Stefano Giuliani
Journal:  Front Pediatr       Date:  2021-12-20       Impact factor: 3.418

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.