Shortening treatment time in robotic radiosurgery using a novel node reduction technique.

Steven van de Water1, Mischa S Hoogeman, Sebastiaan Breedveld, Ben J M Heijmen.   

Abstract

PURPOSE: The fraction duration of robotic radiosurgery treatments can be reduced by generating more time-efficient treatment plans with a reduced number of node positions, beams, and monitor units (MUs). Node positions are preprogramed locations where the robot can position the focal spot of the x-ray beam. As the time needed for the robot to travel between node positions takes up a large part of the treatment time, the aim of this study was to develop and evaluate a node reduction technique in order to reduce the treatment time per fraction for robotic radiosurgery.
METHODS: Node reduction was integrated into the inverse planning algorithm, developed in-house for the robotic radiosurgery modality. It involved repeated inverse optimization, each iteration excluding low-contribution node positions from the planning and resampling new candidate beams from the remaining node positions. Node reduction was performed until the exclusion of a single node position caused a constraint violation, after which the shortest treatment plan was selected retrospectively. Treatment plans were generated with and without node reduction for two lung cases of different complexity, one oropharyngeal case and one prostate case. Plan quality was assessed using the number of node positions, beams and MUs, and the estimated treatment time per fraction. All treatment plans had to fulfill all clinical dose constraints. Extra constraints were added to maintain the low-dose conformality and restrict skin doses during node reduction.
RESULTS: Node reduction resulted in 12 residual node positions, on average (reduction by 77%), at the cost of an increase in the number of beams and total MUs of 28% and 9%, respectively. Overall fraction durations (excluding patient setup) were shortened by 25% (range of 18%-40%), on average. Dose distributions changed only little and dose in low-dose regions was effectively restricted by the additional constraints.
CONCLUSIONS: The fraction duration of robotic radiosurgery treatments can be reduced considerably by node reduction with minimal changes in dosimetrical plan quality. Additional constraints are required to guarantee low-dose conformality and to avoid unacceptable skin dose.

Entities:  

Mesh:

Year:  2011        PMID: 21520851     DOI: 10.1118/1.3549765

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  A phantom study to determine the optimum size of a single collimator for shortening the treatment time in CyberKnife stereotactic radiosurgery of spherical targets.

Authors:  Sudahar Harikrishnaperumal; Gopalakrishna Kurup; Murali Venkatraman; Velmurugan Jagadeesan
Journal:  J Appl Clin Med Phys       Date:  2012-09-06       Impact factor: 2.102

2.  Comparison of multileaf collimator and conventional circular collimator systems in Cyberknife stereotactic radiotherapy.

Authors:  Taro Murai; Yukiko Hattori; Chikao Sugie; Hiromitsu Iwata; Michio Iwabuchi; Yuta Shibamoto
Journal:  J Radiat Res       Date:  2017-09-01       Impact factor: 2.724

3.  Applying pytorch toolkit to plan optimization for circular cone based robotic radiotherapy.

Authors:  Bin Liang; Ran Wei; Jianghu Zhang; Yongbao Li; Tao Yang; Shouping Xu; Ke Zhang; Wenlong Xia; Bin Guo; Bo Liu; Fugen Zhou; Qiuwen Wu; Jianrong Dai
Journal:  Radiat Oncol       Date:  2022-04-20       Impact factor: 4.309

4.  A new plan quality objective function for determining optimal collimator combinations in prostate cancer treatment with stereotactic body radiation therapy using CyberKnife.

Authors:  Maria Varnava; Iori Sumida; Hirokazu Mizuno; Hiroya Shiomi; Osamu Suzuki; Yasuo Yoshioka; Kazuhiko Ogawa
Journal:  PLoS One       Date:  2018-11-27       Impact factor: 3.240

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.