Literature DB >> 21518683

Retrospective image-based gating of intracoronary optical coherence tomography: implications for quantitative analysis.

Kenji Sihan1, Charl Botha, Frits Post, Sebastiaan de Winter, Nieves Gonzalo, Evelyn Regar, Patrick W Serruys, Ronald Hamers, Nico Bruining.   

Abstract

AIMS: Images acquired of coronary vessels during a pullback of time-domain optical coherence tomography (OCT) are influenced by the dynamics of the heart. This study explores the feasibility of applying an in-house developed retrospective image-based gating method for OCT and the influence of catheter dislocation and luminal changes during the cardiac cycle on the outcome of quantitative OCT (QOCT). METHODS AND
RESULTS: The gating method was developed using Matlab (The Mathworks, Natick, MA, USA) and operates in a fully-automatic manner. OCT image data of 20 randomly selected patients, acquired with a commercially available system (Lightlab Imaging, Westford, MA, USA), were pulled from our OCT database for development and validation. Twelve of the 20 datasets could be gated; the other eight pullbacks could not be gated due to a lack of motion induced artefacts. Computations required approximately 30 minutes/dataset. Quantitative comparisons between the gated and the non-gated QOCT results showed significant differences for mean areas and volumes (p <0.001) and mean relative differences of -11% (range -2 up to -20%) for lumen areas (gated) and -13% (range -5 up to -24%) for volumes.
CONCLUSIONS: Retrospective image-based time-domain OCT gating in the presence of motion induced artefacts is feasible. Significant changes in coronary lumen dimensions during the cardiac cycle were observed by OCT and in consequence, quantitative gated OCT analysis showed significant differences compared to non-gated QOCT analyses.

Entities:  

Mesh:

Year:  2011        PMID: 21518683     DOI: 10.4244/EIJV6I9A191

Source DB:  PubMed          Journal:  EuroIntervention        ISSN: 1774-024X            Impact factor:   6.534


  1 in total

1.  The impact of Fourier-Domain optical coherence tomography catheter induced motion artefacts on quantitative measurements of a PLLA-based bioresorbable scaffold.

Authors:  N S van Ditzhuijzen; A Karanasos; N Bruining; M van den Heuvel; O Sorop; J Ligthart; K Witberg; H M Garcia-Garcia; F Zijlstra; D J Duncker; H M M van Beusekom; E Regar
Journal:  Int J Cardiovasc Imaging       Date:  2014-05-16       Impact factor: 2.357

  1 in total