Literature DB >> 21515694

The use of CellaVision competency software for external quality assessment and continuing professional development.

Yuki Horiuchi1, Yoko Tabe, Mayumi Idei, Hans-Inge Bengtsson, Kiyoshi Ishii, Takashi Horii, Kazunori Miyake, Naotake Satoh, Takashi Miida, Akimichi Ohsaka.   

Abstract

AIMS: Quality assessment of blood cell morphological testing, such as white blood cell (WBC) differential and its interpretation, is one of the most important and difficult assignments in haematology laboratories. A monthly survey was performed to assess the possible role of the proficiency testing program produced by CellaVision competency software (CCS) in external quality assessment (EQA) of the clinical laboratories of affiliated university hospitals and the effective utilisation of this program in continuing professional development (CPD).
METHODS: Four monthly proficiency surveys were conducted in collaboration with four clinical laboratories affiliated with the teaching hospitals of Juntendo University of Medicine in Japan.
RESULTS: EQA results by the CCS proficiency testing program revealed a difference of performance levels of WBC differential and morphological interpretation and a discrepancy in the WBC differential criteria among laboratories. With regard to the utilisation of this proficiency program as a tool for CPD, this program successfully improved the performance of the low-scoring laboratories and less experienced individuals.
CONCLUSIONS: The CCS proficiency testing program was useful for the quality assessment of laboratory performance, for education, and for the storage and distribution of cell images to be utilised for further standardisation and education.

Entities:  

Mesh:

Year:  2011        PMID: 21515694     DOI: 10.1136/jcp.2011.089888

Source DB:  PubMed          Journal:  J Clin Pathol        ISSN: 0021-9746            Impact factor:   3.411


  3 in total

1.  Performance of CellaVision DM96 in leukocyte classification.

Authors:  Lik Hang Lee; Adnan Mansoor; Brenda Wood; Heather Nelson; Diane Higa; Christopher Naugler
Journal:  J Pathol Inform       Date:  2013-06-29

2.  Accuracy of the CellaVision DM96 platform for reticulocyte counting.

Authors:  Etienne R Mahe; Diane Higa; Christopher Naugler; Adnan Mansoor; Meer-Taher Shabani-Rad
Journal:  J Pathol Inform       Date:  2014-05-26

3.  Performance of the CellaVision(®) DM96 system for detecting red blood cell morphologic abnormalities.

Authors:  Christopher L Horn; Adnan Mansoor; Brenda Wood; Heather Nelson; Diane Higa; Lik Hang Lee; Christopher Naugler
Journal:  J Pathol Inform       Date:  2015-02-24
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.