Malfunctioning and infected tunneled infusion catheters: over-the-wire catheter exchange versus catheter removal and replacement.

David M Guttmann1, Scott O Trerotola, Timothy W Clark, Mandeep Dagli, Richard D Shlansky-Goldberg, Maxim Itkin, Michael C Soulen, Jeffrey I Mondschein, S William Stavropoulos.   

Abstract

PURPOSE: To compare the safety and effectiveness of over-the-wire catheter exchange (catheter-exchange) with catheter removal and replacement (removal-replacement) at a new site for infected or malfunctioning tunneled infusion catheters.
MATERIALS AND METHODS: Using a quality assurance database, 61 patients with tunneled infusion catheters placed during the period July 2001 to June 2009 were included in this study. Patients receiving hemodialysis catheters were excluded. Catheter-exchange was performed in 25 patients, and same-day removal-replacement was performed in 36 patients. Data collected included demographic information, indication for initial catheter placement and replacement, dwell time for the new catheter, and ultimate fate of the new device. Statistical comparisons between the two cohorts were analyzed using the Kaplan-Meier technique and Fisher exact test.
RESULTS: Catheters exchanged over the wire remained functional without infection for a median of 102 days (range, 2-570 days), whereas catheters removed and replaced were functional for a median 238 days (range, 1-292 days, P = .12). After catheter replacement, there were 11 instances of subsequent infection in the catheter-exchange group and 7 instances in the removal-replacement cohort, accounting for infection rates of 4.4 and 2.3 per 1,000 catheter days (P = .049). Patients in the catheter-exchange group had 3.2 greater odds of infection compared with patients in the removal-replacement group. Five malfunction events occurred in each group, accounting for 2.0 and 1.7 malfunctions per 1,000 catheter days in the catheter-exchange and removal-replacement groups (P = .73).
CONCLUSIONS: Catheter-exchange of tunneled infusion catheters results in a higher infection rate compared with removal-replacement at a new site. The rate of catheter malfunction is not significantly different between the two groups. Catheter-exchange is an alternative for patients with tunneled infusion catheters who have limited venous access, but this technique should not be expanded for use in all patients. Because of the size of this initial study, further investigation is needed to verify the results in a larger sample size.
Copyright © 2011 SIR. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21514518     DOI: 10.1016/j.jvir.2011.01.440

Source DB:  PubMed          Journal:  J Vasc Interv Radiol        ISSN: 1051-0443            Impact factor:   3.464


  2 in total

1.  Guidelines for the prevention of intravascular catheter-related infections: recommendations relevant to interventional radiology for venous catheter placement and maintenance.

Authors:  Donald L Miller; Naomi P O'Grady
Journal:  J Vasc Interv Radiol       Date:  2012-08       Impact factor: 3.464

2.  Radiologic placement of Hickman catheters using intravenous sedation in pediatric patients under 20 kg.

Authors:  Jin Hyeok Kim; Ung Bae Jeon; Joo Yeon Jang; Yong-Woo Kim; Jae-Yeon Hwang; Young Tak Lim; Eu Jeen Yang
Journal:  Medicine (Baltimore)       Date:  2022-02-18       Impact factor: 1.817

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.