Bench study of a new device to display and maintain stable artificial airway cuff pressure.

William R Howard1.   

Abstract

BACKGROUND: Artificial airway cuff pressure should be maintained within a narrow range. Excessive cuff pressure presents a risk of tracheal damage and stenosis. Insufficient cuff pressure adds the risk of secretion leak past the cuff, tidal-volume leakage, and accidental extubation. The available cuff-inflation devices do not address these problems.
OBJECTIVE: In the laboratory I developed and evaluated a new cuff-inflation device that continuously displays the cuff pressure and maintains stable cuff pressure.
METHODS: The cuff-inflation device evaluation included: test the manometer accuracy; compare the displayed pressure to the pressure delivered to the pilot balloon; determine the device's response to cuff-pressure changes with the addition of 5 mL or 10 mL of air after achieving a 30 cm H(2)O baseline; measure the V(T) leak in an intubated artificial trachea by comparing the device results to benchmark measurements; and determine the stability of baseline cuff pressure during routine cuff checks.
RESULTS: The mean ± SD bias and precision of device's display, compared to the calibration analyzer, was 1.3 ± 2.6 cm H(2)O. The pressure delivered by the cuff-inflation device's gas-sampling line to the pilot balloon was equal to the pressure displayed by the cuff-inflation device. With the cuff-inflation device the cuff pressure was unchanged, compared to baseline, after adding 5 mL or 10 mL of air. With 2 current cuff methods, cuff pressure increased to means exceeding 160 cm H(2)O and 300 cm H(2)O, respectively. Compared to the benchmark, the difference in exhaled V(T) mean ± SD bias and precision were: cuff-inflation device 1.4 ± 4.8 mL, and syringe-inflation method 2.4 ± 6.2 mL. Representing a single cuff pressure check, disconnecting the endotracheal-tube pilot balloon from the cuff-inflation device's gas-sampling line and then reconnecting it had no effect on baseline cuff pressure at 2 seconds or 60 seconds.
CONCLUSIONS: The cuff-inflation device demonstrated possible improvements over available cuff-inflation devices and cuff-pressure-control methods.

Entities:  

Mesh:

Year:  2011        PMID: 21513603     DOI: 10.4187/respcare.00961

Source DB:  PubMed          Journal:  Respir Care        ISSN: 0020-1324            Impact factor:   2.258


  1 in total

1.  Decrease in cuff pressure during the measurement procedure: an experimental study.

Authors:  Shota Asai; Asuka Motoyama; Yuri Matsumoto; Hiroyuki Konami; Hideaki Imanaka; Masaji Nishimura
Journal:  J Intensive Care       Date:  2014-06-02
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.