The effect of activity outside the field-of-view on image signal-to-noise ratio for 3D PET with (15)O.

Masanobu Ibaraki1, Shigeki Sugawara, Kazuhiro Nakamura, Fumiko Kinoshita, Toshibumi Kinoshita.   

Abstract

Activity outside the field-of-view (FOV) degrades the count rate performance of 3D PET and consequently reduces signal-to-noise ratios (SNRs) of reconstructed images. The aim of this study was to evaluate a neck-shield installed in a 3D PET scanner for reducing the effect of the outside FOV activity. Specifically, we compared brain PET scans ((15)O(2) and H(2)(15)O) with and without the use of the neck-shield. Image SNRs were directly estimated by a sinogram bootstrap method. The bootstrap analysis showed that the use of the neck-shield improved the SNR by 8% and 19% for H(2)(15)O and (15)O(2), respectively. The SNR improvements were predominantly due to the reduction of the random count rates. Noise equivalent count rate (NECR) analysis provided SNR estimates that were very similar with the bootstrap-based results for H(2)(15)O, but not for (15)O(2). This discrepancy may be due to the fundamental difference between the two methods: the bootstrap method directly calculates the local SNR of reconstructed images, whereas the NECR calculation is based on the whole-gantry count rates, indicating a limitation of the conventional NECR-based method as a tool for assessing the image SNR. Although quantitative parameters, e.g. cerebral blood flow, did not differ when examined with and without the neck-shield, the use of the shield for brain (15)O study is recommended in terms of the image SNR.

Mesh:

Substances:

Year:  2011        PMID: 21508441     DOI: 10.1088/0031-9155/56/10/011

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  7 in total

1.  Reliability of CT perfusion-derived CBF in relation to hemodynamic compromise in patients with cerebrovascular steno-occlusive disease: a comparative study with 15O PET.

Authors:  Masanobu Ibaraki; Tomomi Ohmura; Keisuke Matsubara; Toshibumi Kinoshita
Journal:  J Cereb Blood Flow Metab       Date:  2015-03-11       Impact factor: 6.200

2.  Effects of a novel tungsten-impregnated rubber neck shield on the quality of cerebral images acquired using 15O-labeled gas.

Authors:  Kei Wagatsuma; Keiichi Oda; Kenta Miwa; Motoki Inaji; Muneyuki Sakata; Jun Toyohara; Kiichi Ishiwata; Masayuki Sasaki; Kenji Ishii
Journal:  Radiol Phys Technol       Date:  2017-08-19

3.  Whole-body biodistribution and the influence of body activity on brain kinetic analysis of the 11C-PiB PET scan.

Authors:  Go Akamatsu; Tomoyuki Nishio; Kazuhiko Adachi; Yasuhiko Ikari; Michio Senda
Journal:  Radiol Phys Technol       Date:  2017-09-11

4.  Reconstruction for 3D PET Based on Total Variation Constrained Direct Fourier Method.

Authors:  Haiqing Yu; Zhi Chen; Heye Zhang; Kelvin Kian Loong Wong; Yunmei Chen; Huafeng Liu
Journal:  PLoS One       Date:  2015-09-23       Impact factor: 3.240

5.  Validation of a simplified scatter correction method for 3D brain PET with 15O.

Authors:  Masanobu Ibaraki; Keisuke Matsubara; Kaoru Sato; Tetsuro Mizuta; Toshibumi Kinoshita
Journal:  Ann Nucl Med       Date:  2016-08-17       Impact factor: 2.668

6.  Oxygen-15 labeled CO2, O2, and CO PET in small animals: evaluation using a 3D-mode microPET scanner and impact of reconstruction algorithms.

Authors:  Genki Horitsugi; Tadashi Watabe; Yasukazu Kanai; Hayato Ikeda; Hiroki Kato; Sadahiro Naka; Mana Ishibashi; Keiko Matsunaga; Kayako Isohashi; Eku Shimosegawa; Jun Hatazawa
Journal:  EJNMMI Res       Date:  2017-10-27       Impact factor: 3.138

Review 7.  Quantification of brain oxygen extraction and metabolism with [15O]-gas PET: A technical review in the era of PET/MRI.

Authors:  Audrey P Fan; Hongyu An; Farshad Moradi; Jarrett Rosenberg; Yosuke Ishii; Tadashi Nariai; Hidehiko Okazawa; Greg Zaharchuk
Journal:  Neuroimage       Date:  2020-07-04       Impact factor: 6.556

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.