Literature DB >> 21501065

The incidence of deep brain stimulator hardware infection: the effect of change in antibiotic prophylaxis regimen and review of the literature.

Robin Bhatia1, Arthur Dalton, Mike Richards, Chris Hopkins, Tipu Aziz, Dipankar Nandi.   

Abstract

The complication of hardware infection related to deep brain stimulator implantation (or revision) varies between 0 and 15.2% in the literature. However, no national guidelines exist at present to define an average or acceptable rate of infection associated with, nor the preferred antibiotic prophylaxis required for, this procedure. The aim of this study was to examine the effect of changing the antibiotic prophylaxis regimen used in a single neurosurgical centre on the incidence and outcome of hardware infection. A prospective cohort of 38 patients undergoing deep brain stimulation (DBS) implantation or internal pulse generator (IPG) replacement and receiving perioperative vancomycin (including intravenous gentamicin on induction) and pouch-installed gentamicin, was compared to a historical cohort of 35 patients receiving perioperative cefuroxime in the same unit. The infection rate over 2 years in the prospective group for DBS surgery was 0 compared to 1 (5.6%) in the historical cohort (p = 0.11, χ(2)); the infection rate for IPG replacements was 1(3.6%) in the prospective cohort, versus 3 (17.6%) in the historical (p = 0.44, χ(2)). In this article, we have also systematically reviewed the literature to date and derived an average infection rate of 4.7% (PI 0.9-22%, Random Effects Meta-analysis, Stata) for 35 studies comprising 3550 patients. There is no significant difference in infection rates between DBS procedures that are primarily internalised (n = 9) compared to those in which there is a period of electrode externalisation (n = 23) (p = 0.9, Meta-regression analysis, Stata).

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21501065     DOI: 10.3109/02688697.2011.566384

Source DB:  PubMed          Journal:  Br J Neurosurg        ISSN: 0268-8697            Impact factor:   1.596


  10 in total

1.  Neurosurgery: Psychosurgery guidelines--friction between ideal and reality?

Authors:  Veerle Visser-Vandewalle
Journal:  Nat Rev Neurol       Date:  2014-04-22       Impact factor: 42.937

2.  Deep brain stimulation hardware-related infections: 10-year experience at a single institution.

Authors:  Kingsley O Abode-Iyamah; Hsiu-Yin Chiang; Royce W Woodroffe; Brian Park; Francis J Jareczek; Yasunori Nagahama; Nolan Winslow; Loreen A Herwaldt; Jeremy D W Greenlee
Journal:  J Neurosurg       Date:  2018-03-01       Impact factor: 5.115

3.  Differences in sacral neuromodulation device infection rates based on preoperative antibiotic selection.

Authors:  Allen M Haraway; J Quentin Clemens; Chang He; Cynthia Stroup; Humphrey O Atiemo; Anne P Cameron
Journal:  Int Urogynecol J       Date:  2013-05-22       Impact factor: 2.894

4.  Deep Brain Electrode Externalization and Risk of Infection: A Systematic Review and Meta-Analysis.

Authors:  Alon Kashanian; Pratik Rohatgi; Srinivas Chivukula; Sameer A Sheth; Nader Pouratian
Journal:  Oper Neurosurg (Hagerstown)       Date:  2021-01-13       Impact factor: 2.703

5.  Investigation of deep brain stimulation mechanisms during implantable pulse generator replacement surgery.

Authors:  Brandon D Swan; Warren M Grill; Dennis A Turner
Journal:  Neuromodulation       Date:  2013-10-07

6.  A groove technique for securing an electrode connector on the cranial bone: case analysis of efficacy.

Authors:  Sung-Woo Lee; Myung-Ki Lee; Il Seo; Ho-Sang Kim; Jeong-Ho Kim; Yun-Suk Kim
Journal:  J Korean Neurosurg Soc       Date:  2014-08-31

7.  Surgical site infections after deep brain stimulation surgery: frequency, characteristics and management in a 10-year period.

Authors:  Silje Bjerknes; Inger Marie Skogseid; Terje Sæhle; Espen Dietrichs; Mathias Toft
Journal:  PLoS One       Date:  2014-08-14       Impact factor: 3.240

8.  Update on deep brain stimulation in Parkinson's disease.

Authors:  Daniel Martinez-Ramirez; Wei Hu; Alberto R Bona; Michael S Okun; Aparna Wagle Shukla
Journal:  Transl Neurodegener       Date:  2015-06-27       Impact factor: 8.014

9.  Towards unambiguous reporting of complications related to deep brain stimulation surgery: A retrospective single-center analysis and systematic review of the literature.

Authors:  Katja Engel; Torge Huckhagel; Alessandro Gulberti; Monika Pötter-Nerger; Eik Vettorazzi; Ute Hidding; Chi-Un Choe; Simone Zittel; Hanna Braaß; Peter Ludewig; Miriam Schaper; Kara Krajewski; Christian Oehlwein; Katrin Mittmann; Andreas K Engel; Christian Gerloff; Manfred Westphal; Christian K E Moll; Carsten Buhmann; Johannes A Köppen; Wolfgang Hamel
Journal:  PLoS One       Date:  2018-08-02       Impact factor: 3.240

10.  Experience Reduces Surgical and Hardware-Related Complications of Deep Brain Stimulation Surgery: A Single-Center Study of 181 Patients Operated in Six Years.

Authors:  Mehmet Sorar; Sahin Hanalioglu; Bilge Kocer; Muhammed Taha Eser; Selim Selcuk Comoglu; Hayri Kertmen
Journal:  Parkinsons Dis       Date:  2018-07-22
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.