Micromorphometrical analyses of five different ultrasonic osteotomy devices at the rabbit skull.

Stephan Hollstein1, Eike Hoffmann1, Jürgen Vogel1, Frank Heyroth1, Nora Prochnow1, Peter Maurer1.   

Abstract

OBJECTIVES: The recently introduced ultrasonic osteotome procedure is an alternative to conventional methods of osteotomy. The aim of the present study was to establish the differences between five recently introduced ultrasonic osteotomes and to perform micromorphological and quantitative roughness analyses of osteotomized bone surfaces.
MATERIALS AND METHODS: Fresh, standard-sized bony samples were taken from a rabbit skull using the following ultrasonic osteotomes: the Piezosurgery 3 with insert tip OT7, Piezosurgery Medical with insert tip MT1-10, Piezon Master Surgery with insert tip SL1, VarioSurg with inert tip SG1, and Piezotome 2 with insert tip BS1 II. The required duration of time for each osteotomy was recorded. The prepared surfaces were examined via light microscopy, environmental surface electron microscopy (ESEM), and confocal laser scanning microscopy (CLSM).
RESULTS: All of the investigated piezoelectric osteotomes preserved the anatomical structure of bone. The mean roughness values of the osteotomized bone edge obtained using the investigated piezoelectric osteotomes were as follows: 2.47 μm (Piezosurgery 3), 9.79 μm (Piezosurgery Medical), 4.66 μm (Piezon Master Surgery), 6.38 μm (VarioSurg), and 6.06 μm (Piezotome 2). Significantly higher roughness values were observed when using the Piezosurgery Medical in comparison with those achieved by the Piezosurgery 3 (P<0.0001) and Piezon Master Surgery (P=0.002). Different osteotomy durations were achieved using the different piezoelectric osteotomes: 144 s (Piezosurgery 3), 126 s (Piezosurgery Medical), 142 s (Piezon Master Surgery), 149 s (VarioSurg), and 137 s (Piezotome 2).
CONCLUSIONS: In the present study, micromorphological differences following the use of various ultrasonic devices were clearly identified. According to this study, it can be concluded that the power and the composition of the teeth of the insert tip might impact procedure duration and cutting qualities.
© 2011 John Wiley & Sons A/S.

Entities:  

Mesh:

Year:  2011        PMID: 21488969     DOI: 10.1111/j.1600-0501.2011.02185.x

Source DB:  PubMed          Journal:  Clin Oral Implants Res        ISSN: 0905-7161            Impact factor:   5.977


  3 in total

1.  Corticotomy depth and regional acceleratory phenomenon intensity.

Authors:  Jeremy R Kernitsky; Taisuke Ohira; Dhurata Shosho; June Lim; Abdullah Bamashmous; Serge Dibart
Journal:  Angle Orthod       Date:  2021-03-01       Impact factor: 2.079

2.  Feasibility of piezoelectric endoscopic transsphenoidal craniotomy: a cadaveric study.

Authors:  Peter Valentin Tomazic; Verena Gellner; Wolfgang Koele; Georg Philipp Hammer; Eva Maria Braun; Claus Gerstenberger; Georg Clarici; Etienne Holl; Hannes Braun; Heinz Stammberger; Michael Mokry
Journal:  Biomed Res Int       Date:  2014-02-09       Impact factor: 3.411

Review 3.  Types of devices used in ridge split procedure for alveolar bone expansion: A systematic review.

Authors:  Nayansi Jha; Eun Ha Choi; Nagendra Kumar Kaushik; Jae Jun Ryu
Journal:  PLoS One       Date:  2017-07-21       Impact factor: 3.240

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.