The effect of tissue compression on circular stapler line failure.

Stephan R Myers1, William S Rothermel, Lynn Shaffer.   

Abstract

BACKGROUND: Anastomotic leak after gastrointestinal anastomosis is a well-known and serious complication, yet there is no standardized approach to reliably create an anastomosis with sufficient mechanical properties to consistently avoid an anastomotic leak. The purpose of this study was to evaluate the relationships among combined tissue thickness, tissue compression, closed staple limb length, and mechanical strength of an anastomosis created with a circular stapler, as measured by maximum intraluminal pressure obtained at the time of leakage.
METHODS: Using 27 porcine stomachs and 27 porcine small-intestine segments, we measured tissue thickness and assessed the tissue compression at three different anvil closure distances created by a circular stapling device. Maximum intraluminal pressure was determined by infusing colored water into the porcine materials and increasing the pressure until leakage from the anastomosis occurred.
RESULTS: Tissue compression increased as the closure distance narrowed between the anvil and circular stapler (p < .0001). A tissue compression of ≥6 PSI correlated strongly with a maximum intraluminal pressure of ≥18 mmHg (43% for <6 PSI vs. 90% for ≥6 PSI; p = .02); tissue compression ≥12 PSI was necessary to obtain an acute maximum intraluminal pressure of ≥22 mmHg in 13 of 15 of our samples (p = .04).
CONCLUSIONS: Maximum intraluminal pressure of an anastomosis in this porcine model correlated most strongly with the compression of the tissue involved in the anastomosis. This experimental model provides a framework for constructing a systematic approach to creating an anastomosis with sufficient mechanical strength. However, this study was not intended to establish the upper range of tissue compression beyond which a permanent tissue injury may occur.

Entities:  

Mesh:

Year:  2011        PMID: 21487874     DOI: 10.1007/s00464-011-1667-4

Source DB:  PubMed          Journal:  Surg Endosc        ISSN: 0930-2794            Impact factor:   4.584


  19 in total

1.  Total stapled, total intra-abdominal (TSTI) laparoscopic Roux-en-Y gastric bypass: one leak in 1000 cases.

Authors:  Carlos Carrasquilla; Wayne J English; Paul Esposito; Jennifer Gianos
Journal:  Obes Surg       Date:  2004-05       Impact factor: 4.129

Review 2.  The use of staple-line reinforcement during laparoscopic gastric bypass.

Authors:  Scott A Shikora
Journal:  Obes Surg       Date:  2004 Nov-Dec       Impact factor: 4.129

3.  Roux-en-Y gastric bypass surgery for morbid obesity: evaluation of postoperative extraluminal leaks with upper gastrointestinal series.

Authors:  Laura R Carucci; Mary Ann Turner; Robert C Conklin; Eric J DeMaria; John M Kellum; Harvey J Sugerman
Journal:  Radiology       Date:  2006-01       Impact factor: 11.105

4.  Profiling surgical staplers: effect of staple height, buttress, and overlap on staple line failure.

Authors:  Carlos M Mery; Bilal M Shafi; Gary Binyamin; John M Morton; Michael Gertner
Journal:  Surg Obes Relat Dis       Date:  2008-01-28       Impact factor: 4.734

5.  Laparoscopic repair of gastrointestinal leaks after laparoscopic gastric bypass.

Authors:  Atul K Madan; Brock Lanier; David S Tichansky
Journal:  Am Surg       Date:  2006-07       Impact factor: 0.688

6.  Blood flow in colonic anastomoses. Effect of stapling and suturing.

Authors:  R S Chung
Journal:  Ann Surg       Date:  1987-09       Impact factor: 12.969

7.  Early results after laparoscopic Roux-en-Y gastric bypass: effect of the learning curve.

Authors:  Christopher G Andrew; Wael Hanna; Didier Look; Alexander P H McLean; Nicolas V Christou
Journal:  Can J Surg       Date:  2006-12       Impact factor: 2.089

8.  The role of tissue ischemia in the pathogenesis of anastomotic stricture.

Authors:  R S Chung; D C Hitch; D N Armstrong
Journal:  Surgery       Date:  1988-11       Impact factor: 3.982

9.  Tissue thickness of human stomach measured on excised gastric specimens from obese patients.

Authors:  Hazem Elariny; Hamilton González; Bingshi Wang
Journal:  Surg Technol Int       Date:  2005

Review 10.  Stapled versus handsewn methods for ileocolic anastomoses.

Authors:  P Y G Choy; I P Bissett; J G Docherty; B R Parry; A E H Merrie
Journal:  Cochrane Database Syst Rev       Date:  2007-07-18
View more
  6 in total

1.  A novel single-stapling technique for colorectal anastomosis: a pre-ligation single-stapling technique (L-SST) in a porcine model.

Authors:  Hiroshi Takeyama; Hirofumi Yamamoto; Taishi Hata; Yusuke Takahashi; Masahisa Ohtsuka; Ryoji Nonaka; Akira Inoue; Atsushi Naito; Tae Matsumura; Mamoru Uemura; Junichi Nishimura; Ichiro Takemasa; Tsunekazu Mizushima; Yuichiro Doki; Masaki Mori
Journal:  Surg Endosc       Date:  2014-12-06       Impact factor: 4.584

2.  A Multifunctional Origami Patch for Minimally Invasive Tissue Sealing.

Authors:  Sarah J Wu; Hyunwoo Yuk; Jingjing Wu; Christoph S Nabzdyk; Xuanhe Zhao
Journal:  Adv Mater       Date:  2021-02-01       Impact factor: 30.849

3.  Effectiveness of a new triple-row circular stapler in reducing the risk of colorectal anastomotic leakage: A historical control and propensity score-matched study.

Authors:  Junichi Mazaki; Kenji Katsumata; Tetsuo Ishizaki; Noritoshi Fukushima; Ryutaro Udo; Tomoya Tago; Kenta Kasahara; Hiroshi Kuwabara; Masanobu Enomoto; Yuichi Nagakawa; Akihiko Tsuchida
Journal:  Medicine (Baltimore)       Date:  2022-07-08       Impact factor: 1.817

4.  Compression injury of the circular stapler for gastrointestinal end-to-end anastomosis: preliminary in-vitro study.

Authors:  Gyung Mo Son; Myeong Sook Kwon; Hong-Min Ahn; In Young Lee; Gun Ho Kim; Kyoung Won Nam; JoonWon Lee; Chang-Suk Kong
Journal:  Ann Surg Treat Res       Date:  2020-07-31       Impact factor: 1.859

Review 5.  Surgical stapling device-tissue interactions: what surgeons need to know to improve patient outcomes.

Authors:  Edward Chekan; Richard L Whelan
Journal:  Med Devices (Auckl)       Date:  2014-09-12

6.  Compression automation of circular stapler for preventing compression injury on gastrointestinal anastomosis.

Authors:  Jin-Seok Kim; Sang-Ho Park; Nam-Su Kim; In Young Lee; Hyun Seok Jung; Hong-Min Ahn; Gyung Mo Son; Kwang-Ryul Baek
Journal:  Int J Med Robot       Date:  2022-02-09       Impact factor: 2.483

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.