Microbial contamination and the sterilization/disinfection of surgical guides used in the placement of endosteal implants.

Peter N Smith1, Charles John Palenik, Steven B Blanchard.   

Abstract

PURPOSE: The purpose of this study was twofold: (1) to estimate the level of microbial contamination found on commercial guides (CG) and in-house laboratory guides (LG) prior to use, and (2) to evaluate the antimicrobial potential of disinfectants commonly used in dentistry to decontaminate heat-sensitive surgical guides. Ethylene oxide gas was used as the positive control; sterile water served as a negative control.
MATERIALS AND METHODS: Evaluation of CGs and LGs for bacterial contamination occurred soon after their arrival in the laboratory. Some guides went directly into tubes of trypticase soy broth solution, vortexed and equally divided into two tubes. One tube went into an 80°C bath for 19 minutes, while the other stayed at room temperature. After plating, half of the samples underwent anaerobic incubation. All incubation was for 48 hours at 37°C. Other guides underwent water rinsing or disinfection by various methods for 5 or 15 minutes or ethylene oxide gas sterilization prior to sampling.
RESULTS: Untreated CG specimens showed modest levels of bacterial contamination, with most colonies coming from liquid specimens not exposed to 80°C. LG specimens had more bacteria from both heat-treated and non-heat-treated aliquots. Chlorhexidine gluconate, diluted bleach, and water rinsing were not able to completely eliminate microorganisms on the specimens, but no viable microorganisms were present on specimens treated with gas sterilization or 70% ethanol for 15 minutes.
CONCLUSION: CGs and LGs, on average, had different levels of bacterial contamination prior to disinfection. Water, chlorhexidine gluconate, and diluted bleach were not totally successful in decontaminating the surgical guides, but no growth was found after 15 minutes of immersion in 70% ethanol. Preferably, surgical guides should be submerged in 70% alcohol for a minimum of 15 minutes or undergo sterilization using ethylene oxide gas.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21483880

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  5 in total

1.  Effects of Disinfection and Steam Sterilization on the Mechanical Properties of 3D SLA- and DLP-Printed Surgical Guides for Orthodontic Implant Placement.

Authors:  Silvia Izabella Pop; Mircea Dudescu; Sorin Gheorghe Mihali; Mariana Păcurar; Dana Cristina Bratu
Journal:  Polymers (Basel)       Date:  2022-05-21       Impact factor: 4.967

2.  Efficacy of Debridement for Early Periprosthetic Joint Infection after Hip Arthroplasty.

Authors:  Jong Hoon Kim; Sung Kwang Chun; Yong Cheol Yoon; Devendra Lakhotia; Won Yong Shon
Journal:  Hip Pelvis       Date:  2014-12-31

3.  In Vitro Comparison between Metal Sleeve-Free and Metal Sleeve-Incorporated 3D-Printed Computer-Assisted Implant Surgical Guides.

Authors:  Kyung Chul Oh; June-Sung Shim; Ji-Man Park
Journal:  Materials (Basel)       Date:  2021-01-29       Impact factor: 3.623

4.  Tamoxifen reduces silicone implant capsule formation in a mouse model.

Authors:  Kevin M Blum; Gabriel J M Mirhaidari; Jacob C Zbinden; Christopher K Breuer; Jenny C Barker
Journal:  FASEB Bioadv       Date:  2022-08-16

5.  Printing the Future-Updates in 3D Printing for Surgical Applications.

Authors:  Dekel Shilo; Omri Emodi; Ori Blanc; Dani Noy; Adi Rachmiel
Journal:  Rambam Maimonides Med J       Date:  2018-07-30
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.