Anthony H D Graham1, Chris R Bowen, Susan M Surguy, Jon Robbins, John Taylor.
Abstract
Two new prototype assembly methods have been evaluated for biosensors that combine an integrated circuit (IC) sensor with a culture chamber. The first method uses a poly-ethylene glycol (PEG) mould to mask the IC sensor during application of a room temperature vulcanising (RTV) silicone elastomer used to insulate the bondpads and bondwires. The second method utilises the 'partial encapsulation' service offered by Quik-Pak, USA. Both methods were shown to provide good electrical insulation and demonstrated biocompatibility with the NG108-15 cell line. These methods are particularly useful for the assembly of low-cost ICs with a small (< 4 mm²) sensor area.Entities:
Mesh:
Substances:
Year: 2011 PMID: 21478042 DOI: 10.1016/j.medengphy.2011.03.006