Literature DB >> 21478029

Epithermal neutron beam interference with cardiac pacemakers.

H Koivunoro1, T Serén, H Hyvönen, P Kotiluoto, P Iivonen, I Auterinen, T Seppälä, L Kankaanranta, S Pakarinen, M Tenhunen, S Savolainen.   

Abstract

In this paper, a phantom study was performed to evaluate the effect of an epithermal neutron beam irradiation on the cardiac pacemaker function. Severe malfunction occurred in the pacemakers after substantially lower dose from epithermal neutron irradiation than reported in the fast neutron or photon beams at the same dose rate level. In addition the pacemakers got activated, resulting in nuclides with half-lives from 25 min to 115 d. We suggest that BNCT should be administrated only after removal of the pacemaker from the vicinity of the tumor. Copyright Â
© 2011 Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21478029     DOI: 10.1016/j.apradiso.2011.03.028

Source DB:  PubMed          Journal:  Appl Radiat Isot        ISSN: 0969-8043            Impact factor:   1.513


  1 in total

1.  Spot-scanning proton therapy for targets with adjacent cardiac implantable electronic devices - Strategies for breast and head & neck cancer.

Authors:  Line Bjerregaard Stick; Peter Magnus Trock Lægdsmand; Henrik Laurits Bjerre; Morten Høyer; Kenneth Jensen; Maria Fuglsang Jensen; Mads Brix Kronborg; Birgitte Vrou Offersen; Camilla Jensenius Skovhus Kronborg