The impact of capitellar arthroplasty on elbow contact mechanics: Implications for implant design.

M T Sabo1, H Shannon, J Ng, L M Ferreira, J A Johnson, G J W King.   

Abstract

BACKGROUND: Radiocapitellar arthroplasty is indicated for capitellar deficiency. Although current implants employ a spherical capitellar surface, the capitellum is elliptical. This has implications for congruency and wear in capitellar arthroplasty. Our objective was to evaluate the contact mechanics of radiocapitellar arthroplasty. We hypothesized that capitellar replacement would decrease joint contact area relative to the native articulation; and an anatomical implant design would replicate more normal contact morphology than a spherical implant.
METHODS: Eight paired humeri and radii were potted in a custom jig. A compressive load of 85N was applied with the articulation flexed at 45° in neutral forearm rotation. Joint casts were made and the articular contact area and shape were quantified. Anatomical and spherical capitellar implants were tested against the native radial head (hemiarthroplasty).
FINDINGS: The contact areas for the anatomical and spherical hemiarthroplasties were 59 and 51% of the native articulation (P<0.005), while the unicompartmental arthroplasties' contact areas were 84 and 89% (anatomical, spherical) of the native articulation (P<0.01). No implant was superior in recreating the native contact shape.
INTERPRETATION: Placement of any capitellar implant resulted in a large decrease in contact area when articulating with a native radial head. This suggests that the radial head cartilage would see a marked increase in contact pressure relative to the native articulation. The unicompartmental arthroplasties demonstrated an even larger reduction in contact area, raising concern about accelerated surface wear. Further investigation needs to correlate these contact mechanics to cartilage wear and implant longevity.
Copyright © 2011 Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21474220     DOI: 10.1016/j.clinbiomech.2011.01.007

Source DB:  PubMed          Journal:  Clin Biomech (Bristol, Avon)        ISSN: 0268-0033            Impact factor:   2.063


  3 in total

1.  Capitellum Fracture Fragment Excision: a Case Series.

Authors:  Matthew R Garner; Patrick C Schottel; Robert N Hotchkiss; Aaron Daluiski; Dean G Lorich
Journal:  HSS J       Date:  2015-08-01

2.  Radiocapitellar prosthetic arthroplasty in traumatic and post-traumatic complex lesions of the elbow.

Authors:  Prospero Bigazzi; Marco Biondi; Massimo Ceruso
Journal:  Eur J Orthop Surg Traumatol       Date:  2016-08-20

3.  From radial head to radiocapitellar to total elbow replacement: A case report.

Authors:  Michael Hackl; Klaus Josef Burkhart; Kilian Wegmann; Boris Hollinger; Sven Lichtenberg; Lars Peter Müller
Journal:  Int J Surg Case Rep       Date:  2015-08-14
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.