Literature DB >> 21469968

Targeting accuracy of CT-guided stereotaxy for radiofrequency ablation of liver tumours.

Gerlig Widmann1, Peter Schullian, Marion Haidu, Martin Fasser, Reto Bale.   

Abstract

The targeting accuracy during CT-guided stereotactic radiofrequency ablation (SRFA) of liver tumours was evaluated in a clinical study. Patients under general anaesthesia were immobilized using a vacuum cushion and respiratory motion control was based on temporary disconnections of the endotracheal tube. An optical-based navigation system was used for 3D trajectory planning and needle placement via a stereotactic aiming device. A control CT with the needles in place was fused with the planning CT for accuracy evaluation. Sub-analysis was performed for "location" (liver segments II-VIII), "approach" (intercostal or transabdominal), "properties" (clear parenchymal, subcapsular, subphrenic, fat, and subphrenic plus fat), and "path length". In 20 patients with 35 liver lesions, a total of 145 needles were placed with mean (±SD) lateral errors of 3.6 ± 2.5 mm at the needle tip, angular errors of 1.3° ± 1.2°, and longitudinal errors at the needle tip of -7.4 ± 6.2 mm. No puncture-related complications were noted. No significant differences of angular errors between segments, approach and properties were recorded. Long paths correlated with smaller angular errors. CT-guided stereotaxy can be considered safe and provided precise multiple needle placement for SRFA of liver lesions at arbitrary trajectories in various segments and locations.

Entities:  

Mesh:

Year:  2011        PMID: 21469968     DOI: 10.3109/13645706.2010.533923

Source DB:  PubMed          Journal:  Minim Invasive Ther Allied Technol        ISSN: 1364-5706            Impact factor:   2.442


  6 in total

1.  Patient-specific needle guidance templates drilled intraprocedurally for image guided intervention: feasibility study in swine.

Authors:  Neil Glossop; Reto Bale; Sheng Xu; William F Pritchard; John W Karanian; Bradford J Wood
Journal:  Int J Comput Assist Radiol Surg       Date:  2022-09-29       Impact factor: 3.421

2.  Simulated accuracy assessment of small footprint body-mounted probe alignment device for MRI-guided cryotherapy of abdominal lesions.

Authors:  Naoyuki Shono; Brian Ninni; Franklin King; Takahisa Kato; Junichi Tokuda; Takahiro Fujimoto; Kemal Tuncali; Nobuhiko Hata
Journal:  Med Phys       Date:  2020-03-28       Impact factor: 4.071

3.  Safety and efficacy of stereotactic radiofrequency ablation for very large (≥8 cm) primary and metastatic liver tumors.

Authors:  Peter Schullian; Edward W Johnston; Daniel Putzer; Gernot Eberle; Gregor Laimer; Reto Bale
Journal:  Sci Rep       Date:  2020-01-31       Impact factor: 4.379

4.  Feasibility, safety, and long-term efficacy of stereotactic radiofrequency ablation for tumors adjacent to the diaphragm in the hepatic dome: a case-control study.

Authors:  Peter Schullian; Daniel Putzer; Gregor Laimer; Elliot Levy; Reto Bale
Journal:  Eur Radiol       Date:  2019-09-05       Impact factor: 5.315

5.  Primary efficacy of percutaneous microwave ablation of malignant liver tumors: comparison of stereotactic and conventional manual guidance.

Authors:  Jan Schaible; Lukas Lürken; Philipp Wiggermann; Niklas Verloh; Ingo Einspieler; Florian Zeman; Andreas G Schreyer; Reto Bale; Christian Stroszczynski; Lukas Beyer
Journal:  Sci Rep       Date:  2020-11-02       Impact factor: 4.379

6.  Stereotactic and Robotic Minimally Invasive Thermal Ablation of Malignant Liver Tumors: A Systematic Review and Meta-Analysis.

Authors:  Pascale Tinguely; Iwan Paolucci; Simeon J S Ruiter; Stefan Weber; Koert P de Jong; Daniel Candinas; Jacob Freedman; Jennie Engstrand
Journal:  Front Oncol       Date:  2021-09-23       Impact factor: 6.244

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.