Pressure generated by syringes: implications for hydrodissection and injection of dense connective tissue lesions.

W A P Hayward1, L J Haseler, L G Kettwich, A A Michael, W l Sibbitt, A D Bankhurst.   

Abstract

OBJECTIVE: Hydrodissection and high-pressure injection are important for the treatment of dense connective tissue lesions including rheumatoid nodules, Dupuytren's contracture, and trigger finger. The present study determined the optimal syringes for high-pressure injection of dense connective tissue lesions.
METHODS: Different sizes (1, 3, 5, 10, 20, and 60 mL) of a mechanical syringe (reciprocating procedure device) with a luer-lock fitting were studied. Twenty operators generated maximum pressure with each mechanical syringe size, and pressure was measured in pounds per square inch (psi). Subsequently, 223 dense connective tissue lesions were injected with different sizes of syringes (1, 3, or 10 mL). Outcomes included (i) successful intralesional injection and (ii) clinical response at 2 weeks.
RESULTS: Smaller syringes generated significantly more injection pressure than did larger syringes: 1 mL (363 ± 197 psi), 3 mL (177 ± 96 psi), 5 mL (73 ± 40 psi), 10 mL (53 ± 29 psi), 20 mL (32 ± 18 psi), and 60 mL (19 ± 12 psi). Similarly, smaller syringes were superior to larger syringes for intralesional injection success: 10 mL: 34% (15/44) vs. 1 mL: 100% (70/70) (p < 0.001) and 3 mL: 91% (99/109) (p < 0.001).
CONCLUSION: Smaller syringes (≤ 3 mL) are superior to larger syringes (≥ 5 mL) for successful hydrodissection and high-pressure intralesional injection of dense connective tissue lesions.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21469942     DOI: 10.3109/03009742.2011.560892

Source DB:  PubMed          Journal:  Scand J Rheumatol        ISSN: 0300-9742            Impact factor:   3.641


  6 in total

1.  Outcomes and Complications of Corticosteroid Injection of Rheumatoid Nodules.

Authors:  Angie Ariza-Hutchinson; Rosemina A Patel; N Suzanne Emil; Maheswari Muruganandam; Sharon E Nunez; Matthew K McElwee; Frank X O'Sullivan; Roderick A Fields; William A Hayward; Luke J Haseler; Wilmer L Sibbitt
Journal:  J Clin Aesthet Dermatol       Date:  2022-08

2.  The flushing procedure in nursing practices: A cross-sectional study with Portuguese and Brazilian nurses.

Authors:  Pedro Parreira; Ricardo Vicente; Rafael A Bernardes; Liliana B Sousa; Beatriz Serambeque; Paulo Costa; Luciene M Braga; Lisete Mónico; Anabela Salgueiro-Oliveira
Journal:  Heliyon       Date:  2020-08-08

3.  Rupture of an epidural filter connector during bolus administration of local anesthetic: a case report.

Authors:  Daniel A Nahrwold; Aaron R Muncey; Nasrin N Aldawoodi; Raymond M Evans; Jamie P Hoffman
Journal:  BMC Anesthesiol       Date:  2021-05-12       Impact factor: 2.217

4.  Pressure-Volume Relationships in the Spinal Canal and Potential Neurological Complications After Epidural Fluid Injections.

Authors:  Hemmo Bosscher
Journal:  Front Pain Res (Lausanne)       Date:  2022-07-07

5.  Conducting polymer-based granular hydrogels for injectable 3D cell scaffolds.

Authors:  Vivian Rachel Feig; Sruthi Santhanam; Kelly Wu McConnell; Kathy Liu; Matine Azadian; Lucia Giulia Brunel; Zhuojun Huang; Helen Tran; Paul M George; Zhenan Bao
Journal:  Adv Mater Technol       Date:  2021-04-25

6.  A new BiofilmChip device for testing biofilm formation and antibiotic susceptibility.

Authors:  Núria Blanco-Cabra; Maria José López-Martínez; Betsy Verónica Arévalo-Jaimes; María Teresa Martin-Gómez; Josep Samitier; Eduard Torrents
Journal:  NPJ Biofilms Microbiomes       Date:  2021-08-03       Impact factor: 7.290

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.