Literature DB >> 21459603

Beating-heart implantation of adjustable length mitral valve chordae: acute and chronic experience in an animal model.

Francesco Maisano1, Micaela Cioni, Joerg Seeburger, Volkmar Falk, Friedrich Wilhelm Mohr, Michael J Mack, Ottavio Alfieri, Hugo Vanermen.   

Abstract

OBJECTIVE: This study aimed to determine the acute and chronic performance of a new system designed to conduct beating-heart implantation and off-pump adjustment of neochordal length.
METHODS: In 14 adult sheep (group A) selected to undergo beating-heart cardiopulmonary bypass, the left atrium was opened through a left thoracotomy. Two or more primary chordae in the A2 region were severed to produce a model of a flail leaflet. A chordal adjustment mechanism (V-Chordal, Valtech Cardio Ltd., Or-Yehuda, Israel) was affixed to the head of the papillary muscle. The system includes two adjustable neochordae. The distal end of the neochordae was sutured to the flail segment without estimating the appropriate length. The neochordal length was adjusted off-pump under real-time echo-guidance. The adjustment tool was removed and the atriotomy was closed with a purse-string suture. Control animals (group B, n=4) were implanted with the conventional neochordae. Animals in both groups were sacrificed 3 months after the procedure.
RESULTS: In both groups, prior to repair, mitral regurgitation (MR) was severe in all animals. In group A, following adjustment of neochordae, MR was absent in all animals, with the exception of two animals that had residual 2+ MR irresponsive to neochordae adjustments. In group B, MR was 2+ in two of the four animals following repair. At 3 months, mitral competence was stable in all animals. At necropsy, normal healing of the papillary head and leaflet was observed in both the groups.
CONCLUSIONS: The V-Chordal system simplifies the process of neochordal implantation and precise off-pump adjustment of the neochordal length to correct MR occurring due to a flail leaflet. This technology may improve the technical feasibility for adoption of chordal repair during open or minimally invasive surgical procedures.
Copyright © 2011 European Association for Cardio-Thoracic Surgery. Published by Elsevier B.V. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21459603     DOI: 10.1016/j.ejcts.2011.01.021

Source DB:  PubMed          Journal:  Eur J Cardiothorac Surg        ISSN: 1010-7940            Impact factor:   4.191


  9 in total

1.  Mitral valve repair: historical evolution and what lies ahead.

Authors:  Ugo Filippo Tesler
Journal:  Tex Heart Inst J       Date:  2012

Review 2.  Evidence for the role of gonadotropin hormones in the development of Alzheimer disease.

Authors:  G Casadesus; C S Atwood; X Zhu; A W Hartzler; K M Webber; G Perry; R L Bowen; M A Smith
Journal:  Cell Mol Life Sci       Date:  2005-02       Impact factor: 9.261

Review 3.  Devices for mitral valve repair.

Authors:  Paolo Denti; Francesco Maisano; Ottavio Alfieri
Journal:  J Cardiovasc Transl Res       Date:  2014-01-23       Impact factor: 4.132

Review 4.  Treatment and management of mitral regurgitation.

Authors:  Michele De Bonis; Francesco Maisano; Giovanni La Canna; Ottavio Alfieri
Journal:  Nat Rev Cardiol       Date:  2011-11-22       Impact factor: 32.419

Review 5.  Interventional approaches to mitral regurgitation. What's in the pipeline?

Authors:  V Mauri; S Baldus; V Rudolph
Journal:  Herz       Date:  2016-02       Impact factor: 1.443

Review 6.  Artificial chordae for degenerative mitral valve disease: critical analysis of current techniques.

Authors:  Michael Ibrahim; Christopher Rao; Thanos Athanasiou
Journal:  Interact Cardiovasc Thorac Surg       Date:  2012-09-07

7.  Mitral transcatheter technologies.

Authors:  Francesco Maisano; Nicola Buzzatti; Maurizio Taramasso; Ottavio Alfieri
Journal:  Rambam Maimonides Med J       Date:  2013-07-25

8.  Mitral chordae myxoma-chordae replacement with a premeasured gore-tex loop using a minimally invasive video-assisted approach.

Authors:  Masatoshi Hata; Jan F Gummert; Jochen Börgermann; Kavous Hakim-Meibodi
Journal:  J Cardiothorac Surg       Date:  2013-12-11       Impact factor: 1.637

9.  ChoRe: A device for trans-catheter chordae tendineae repair.

Authors:  Costanza Culmone; Awaz Ali; Marta Scali; Arianna Menciassi; Paul Breedveld
Journal:  Proc Inst Mech Eng H       Date:  2019-05-08       Impact factor: 1.617

  9 in total