Literature DB >> 21458995

A new device for in situ static and dynamic calibration of force platforms.

Hong-Jung Hsieh1, Tung-Wu Lu, Sheng-Chang Chen, Chia-Min Chang, Chinghua Hung.   

Abstract

In human motion analysis, in situ calibration of the force plate is necessary to improve the accuracy of the measured ground reaction force (GRF) and center of pressure (COP). Few existing devices are capable of both static and dynamic calibration of the usually non-linear GRF and COP errors, while are also easy to move and/or set up without damaging the building. The current study developed a small device (160 cm × 88 cm × 43 cm) with a mass of 50 kg, equipped with auxiliary wheels and fixing suction pads for rapid deployment and easy set-up. A PC-based controller enabled quick movement and accurate positioning of the applied force to the calibration point. Static calibration at 100 validation points and dynamic calibration of a force plate were performed using the device. After correction by an artificial neural network (ANN) trained with the static data from another 121 points, the mean errors for the GRF were all reduced from a maximum of 0.64% to less than 0.01%, while those for the COP were all reduced from a maximum of about 1.37 mm to less than 0.04 mm. For dynamic calibration, the mean errors for the GRF were reduced from a maximum of 0.46% to less than 0.28%, while those for the COP were reduced from a maximum of 0.95 mm to less than 0.11 mm. The results suggest that the calibration device with the ANN method will be useful for obtaining more accurate GRF and COP measurements in human motion analysis.
Copyright © 2011 Elsevier B.V. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21458995     DOI: 10.1016/j.gaitpost.2011.03.005

Source DB:  PubMed          Journal:  Gait Posture        ISSN: 0966-6362            Impact factor:   2.840


  4 in total

1.  Dynamic Outlier Detection in the Calibration by Comparison Method Applied to Strain Gauge Weight Sensors.

Authors:  Wojciech Walendziuk
Journal:  Sensors (Basel)       Date:  2018-11-30       Impact factor: 3.576

2.  Tendon release reduced joint stiffness with unaltered leg stiffness during gait in spastic diplegic cerebral palsy.

Authors:  Chien-Chung Kuo; Hsing-Po Huang; Ting-Ming Wang; Shih-Wun Hong; Li-Wei Hung; Ken N Kuo; Tung-Wu Lu
Journal:  PLoS One       Date:  2021-01-15       Impact factor: 3.240

3.  Avascular Necrosis of the Hip Compromises Gait Balance Control in Female Juveniles With Unilateral Developmental Dysplasia Treated in Toddlerhood.

Authors:  Wei-Chun Lee; Pei-An Lee; Tsan-Yang Chen; Yu-Ting Chen; Kuan-Wen Wu; Yu-Lin Tsai; Ting-Ming Wang; Tung-Wu Lu
Journal:  Front Bioeng Biotechnol       Date:  2022-03-24

4.  Synergistic multi-joint kinematic strategies to reduce tripping risks during obstacle-crossing in older long-term Tai-Chi Chuan practitioners.

Authors:  Hsing-Po Huang; Chien-Chung Kuo; Shiuan-Huei Lu; Sheng-Chang Chen; Tsung-Jung Ho; Tung-Wu Lu
Journal:  Front Aging Neurosci       Date:  2022-09-29       Impact factor: 5.702

  4 in total