BNCT treatment planning of recurrent head-and-neck cancer using THORplan.

Hui-Ting Yu1, Yen-Wan Hsueh Liu, Tzung-Yi Lin, Ling-Wei Wang.   

Abstract

A cooperation program on Boron Neutron Capture Therapy (BNCT) between National Tsing Hua University (NTHU) and Taipei Veterans General Hospital (TPEVGH) was established in 2008. Clinical trial of recurrent head-and-neck cancer is the goal of the program. In this study, treatment plannings of two head-and-neck cancer cases are performed using treatment planning system THORplan developed at NTHU of Taiwan. The patients are assumed to be irradiated under current THOR epithermal neutron beam. The prescription dose is 20 Gy-Eq for at least 80% of tumor volume. The irradiation time to reach the target tumor dose can be kept within 1h. The skin dose is within the limiting dose of 11 Gy-Eq. The spinal cord dose is well within the limiting dose of 10 Gy-Eq. The use of an extension collimator for easier patient positioning is helpful in reducing the dose of eye lens to within the dose limit of 5 Gy-Eq. The irradiation time, however, will increase slightly due to the increase of source-to-tumor distance. The CPU time for treatment planning calculation is ~10 h. With the use of user friendly treatment planning system THORplan, dose planning for BNCT at THOR can be easily performed. Copyright Â
© 2011 Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21458281     DOI: 10.1016/j.apradiso.2011.02.041

Source DB:  PubMed          Journal:  Appl Radiat Isot        ISSN: 0969-8043            Impact factor:   1.513


  3 in total

Review 1.  Physics of epi-thermal boron neutron capture therapy (epi-thermal BNCT).

Authors:  Ryoichi Seki; Yushi Wakisaka; Nami Morimoto; Masaaki Takashina; Masahiko Koizumi; Hiroshi Toki; Mitsuhiro Fukuda
Journal:  Radiol Phys Technol       Date:  2017-11-20

Review 2.  Clinical trials for treating recurrent head and neck cancer with boron neutron capture therapy using the Tsing-Hua Open Pool Reactor.

Authors:  Ling-Wei Wang; Yen-Wan Hsueh Liu; Fong-In Chou; Shiang-Huei Jiang
Journal:  Cancer Commun (Lond)       Date:  2018-06-19

3.  A comparison of dose distributions in gross tumor volume between boron neutron capture therapy alone and combined boron neutron capture therapy plus intensity modulation radiation therapy for head and neck cancer.

Authors:  Jia-Cheng Lee; Keh-Shih Chuang; Yen-Wan Hsueh Liu; Tzung-Yi Lin; Yi-Chiao Teng; Ling-Wei Wang
Journal:  PLoS One       Date:  2019-04-04       Impact factor: 3.240

  3 in total