| Literature DB >> 21458090 |
Barbara Dobler1, Christian Groeger, Marius Treutwein, Judith Alvarez-Moret, Thomas Goetzfried, Karin Weidner, Petra Haertl, Oliver Koelbl.
Abstract
Methods and results for commissioning of the complete VMAT delivery chain are presented for the combination of Nucletron's Oncentra MasterPlan® v3.3 with Elekta's Mosaiq® v1.6 and SynergyS® linac. VMAT specific linac commissioning included determination of the size of the minimal dynamic leaf gap. Dosimetric validation of the complete treatment chain was performed using a 2D-ionization-chamber-array and showed excellent dosimetric results.Mesh:
Year: 2011 PMID: 21458090 DOI: 10.1016/j.radonc.2011.01.024
Source DB: PubMed Journal: Radiother Oncol ISSN: 0167-8140 Impact factor: 6.280