Literature DB >> 21456914

Shunt failure due to intracranial migration of BioGlide ventricular catheters.

Hsin-Hung Chen1, Jay Riva-Cambrin, Douglas L Brockmeyer, Marion L Walker, John R W Kestle.   

Abstract

OBJECT: In late 2008, the authors recognized a new type of ventriculoperitoneal shunt failure specific to the Bio-Glide Snap Shunt ventricular catheters. This prompted a retrospective review of the patient cohort and resulted in a recall by the FDA in the US.
METHODS: After the index cases were identified, the FDA was notified by the hospital, leading to a recall of the product. Hospital operative logs were used to identify patients in whom the affected products were used. A letter describing the risk was sent to all patients offering a free screening CT scan to look for disconnection. A call center was established to respond to patient questions, and an informational video was made available on the hospital website. The authors reviewed the records of the index cases and other cases subsequently identified.
RESULTS: Seven index cases and an additional 16 cases of disconnection were identified in the 466 patients in whom a BioGlide Snap Shunt ventricular catheter had been implanted. Mean time to disconnection was 2.7 years (range 4 days-5.8 years). Computed tomography slices in the plane of the catheter helped visualize disconnections. Retrieval was difficult, and in 5 patients the disconnected catheter was not removable. Three catheters were completely within the ventricle. At presentation, 4 children suffered from severe neurological deficits, including one who died as a result of the shunt malfunction.
CONCLUSIONS: BioGlide snap-design ventricular catheters are prone to disconnection. Continued vigilance and specific imaging are important. Catheter removal after disconnection may be difficult. Elective removal prior to disconnection in asymptomatic children has not been performed.

Entities:  

Mesh:

Year:  2011        PMID: 21456914     DOI: 10.3171/2011.1.PEDS10389

Source DB:  PubMed          Journal:  J Neurosurg Pediatr        ISSN: 1933-0707            Impact factor:   2.375


  3 in total

Review 1.  Cerebrospinal Fluid Shunting Complications in Children.

Authors:  Brian W Hanak; Robert H Bonow; Carolyn A Harris; Samuel R Browd
Journal:  Pediatr Neurosurg       Date:  2017-03-02       Impact factor: 1.162

2.  Evaluation of an in vivo model for ventricular shunt infection: a pilot study using a novel antimicrobial-loaded polymer.

Authors:  Rajiv R Iyer; Noah Gorelick; Karen Carroll; Ari M Blitz; Sarah Beck; Caroline M Garrett; Audrey Monroe; Betty Tyler; Sean T Zuckerman; Jeffrey R Capadona; Horst A von Recum; Mark G Luciano
Journal:  J Neurosurg       Date:  2018-08-03       Impact factor: 5.115

3.  The Shunt Slippage: A Complication of Pumping Test.

Authors:  El Kim
Journal:  Korean J Neurotrauma       Date:  2022-02-10
  3 in total