Literature DB >> 21456790

Mass determination and sensitivity based on resonance frequency changes of the higher flexural modes of cantilever sensors.

John D Parkin1, Georg Hähner.   

Abstract

Micro- and nanocantilevers are increasingly employed as mass sensors. Most studies consider the first flexural mode and adsorbed masses that are either discretely attached or homogeneously distributed along the entire length of the cantilever. We derive general expressions that allow for the determination of the total attached mass with any mass distribution along the cantilever length and all flexural modes. The expressions are valid for all cantilevers whose flexural deflection can be described by a one-dimensional function. This approach includes the most common types of microcantilevers, namely, rectangular, picket, and V-shaped. The theoretical results are compared with experimental data up to the fourth flexural mode obtained from thermal noise spectra of rectangular and V-shaped cantilevers.

Year:  2011        PMID: 21456790     DOI: 10.1063/1.3563724

Source DB:  PubMed          Journal:  Rev Sci Instrum        ISSN: 0034-6748            Impact factor:   1.523