Continuous cardiac output monitoring with an uncalibrated pulse contour method in patients supported with mechanical pulsatile assist device.

Sabino Scolletta1, Fabio Miraldi, Salvatore Mario Romano, Luigi Muzzi.   

Abstract

We evaluated the accuracy of an uncalibrated pulse contour method called Pressure Recording Analytical Method (PRAM) compared with continuous thermodilution for cardiac output (CO) monitoring in patients implanted with a pulsatile left ventricular assist device (LVAD). Twelve adult patients implanted with the HeartMate I-XVE device were studied. CO was simultaneously evaluated by PRAM and by continuous thermodilution. Blood flow values displayed by the LVAD's console were also used for the comparison. Bland-Altman and linear regression analyses were applied. A total of 72 CO measurements (range 3.8-6.7 l/min) were obtained. Mean CO was 5.23±0.70 l/min for the 'hot' pulmonary thermodilution (ThD-CCO) method, 5.28±0.63 l/min for PRAM and 4.83±0.67 l/min for LVAD-CO. A high correlation (r=0.90), a good agreement (mean bias -0.04 l/min, precision ±0.38 l/min) and a low percentage of error (7.3%) were observed between PRAM-CO and ThD-CCO. A good correlation was found between LVAD-CO and either ThD-CCO (r=0.88) or PRAM-CO (r=0.86), but an overestimation of 10% was observed for both PRAM-CO (mean bias -0.44 l/min) and ThD-CCO (mean bias -0.40 l/min). Our results demonstrated good agreements between PRAM-CO, ThD-CCO and LVAD-CO. PRAM derives CO from a peripheral artery without calibration and may be a complementary tool in the hemodynamic assessment of patients supported with a VAD.

Entities:  

Mesh:

Year:  2011        PMID: 21454314     DOI: 10.1510/icvts.2010.264234

Source DB:  PubMed          Journal:  Interact Cardiovasc Thorac Surg        ISSN: 1569-9285


  3 in total

1.  Monitoring Cardiac Output and Transesophageal Echocardiography during Removal of a Ventricular Assist Device.

Authors:  Aslı Demir; Ümit Karadeniz; Bahar Aydınlı; Murat Taş; Özcan Erdemli
Journal:  Turk J Anaesthesiol Reanim       Date:  2013-05-23

2.  A preliminary study evaluating cardiac output measurement using Pressure Recording Analytical Method (PRAM) in anaesthetized dogs.

Authors:  Angela Briganti; Flavia Evangelista; Paola Centonze; Annaliso Rizzo; Francesco Bentivegna; Antonio Crovace; Francesco Staffieri
Journal:  BMC Vet Res       Date:  2018-03-06       Impact factor: 2.741

3.  Noninvasive evaluation of the hemodynamic status in patients after heart transplantation or left ventricular assist device implantation.

Authors:  Sebastian Roth; Henrik Fox; René M'Pembele; Michiel Morshuis; Giovanna Lurati Buse; Markus W Hollmann; Ragnar Huhn; Thomas Bitter
Journal:  PLoS One       Date:  2022-10-14       Impact factor: 3.752

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.