Aperture modulated, translating bed total body irradiation.

Amjad Hussain1, Jose Eduardo Villarreal-Barajas, Peter Dunscombe, Derek W Brown.   

Abstract

PURPOSE: Total body irradiation (TBI) techniques aim to deliver a uniform radiation dose to a patient with an irregular body contour and a heterogeneous density distribution to within +/-10% of the prescribed dose. In the current article, the authors present a novel, aperture modulated, translating bed TBI (AMTBI) technique that produces a high degree of dose uniformity throughout the entire patient.
METHODS: The radiation beam is dynamically shaped in two dimensions using a multileaf collimator (MLC). The irregular surface compensation algorithm in the Eclipse treatment planning system is used for fluence optimization, which is performed based on penetration depth and internal inhomogeneities. Two optimal fluence maps (AP and PA) are generated and beam apertures are created to deliver these optimal fluences. During treatment, the patient/phantom is translated on a motorized bed close to the floor (source to bed distance: 204.5 cm) under a stationary radiation beam with 0 degree gantry angle. The bed motion and dynamic beam apertures are synchronized.
RESULTS: The AMTBI technique produces a more homogeneous dose distribution than fixed open beam translating bed TBI. In phantom studies, the dose deviation along the midline is reduced from 10% to less than 5% of the prescribed dose in the longitudinal direction. Dose to the lung is reduced by more than 15% compared to the unshielded fixed open beam technique. At the lateral body edges, the dose received from the open beam technique was 20% higher than that prescribed at umbilicus midplane. With AMTBI the dose deviation in this same region is reduced to less than 3% of the prescribed dose. Validation of the technique was performed using thermoluminescent dosimeters in a Rando phantom. Agreement between calculation and measurement was better than 3% in all cases.
CONCLUSIONS: A novel, translating bed, aperture modulated TBI technique that employs dynamically shaped MLC defined beams is shown to improve dose uniformity in three dimensions. In comparison with the fixed open beam TBI technique, homogeneity of dose distribution is greatly improved.

Entities:  

Mesh:

Year:  2011        PMID: 21452729     DOI: 10.1118/1.3534196

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  7 in total

1.  Total body irradiation with step translation and dynamic field matching.

Authors:  Ho-Hsing Chen; Jay Wu; Keh-Shih Chuang; Jia-Fu Lin; Jia-Cheng Lee; Jin-Ching Lin
Journal:  Biomed Res Int       Date:  2013-07-01       Impact factor: 3.411

2.  Total body irradiation dose optimization based on radiological depth.

Authors:  Amjad Hussain; Peter Dunscombe; J Eduardo Villarreal-Barajas; Derek Brown
Journal:  J Appl Clin Med Phys       Date:  2012-05-10       Impact factor: 2.102

3.  Optimized Conformal Total Body Irradiation methods with Helical TomoTherapy and Elekta VMAT: Implementation, Imaging, Planning and Dose Delivery for Pediatric Patients.

Authors:  Anna Anzorovna Loginova; Diana Anatolievna Tovmasian; Anastasiya Olegovna Lisovskaya; Daria Alexeevna Kobyzeva; Michael Alexandrovich Maschan; Alexander Petrovich Chernyaev; Oleg Borisovich Egorov; Alexey Vladimirovich Nechesnyuk
Journal:  Front Oncol       Date:  2022-03-10       Impact factor: 6.244

4.  A Monte Carlo evaluation of beam characteristics for total body irradiation at extended treatment distances.

Authors:  Roumiana Chakarova; Marcus Krantz
Journal:  J Appl Clin Med Phys       Date:  2014-05-08       Impact factor: 2.102

5.  Going the distance: validation of Acuros and AAA at an extended SSD of 400 cm.

Authors:  Narottam Lamichhane; Vivek N Patel; Matthew T Studenski
Journal:  J Appl Clin Med Phys       Date:  2016-03-08       Impact factor: 2.102

6.  Translating bed total body irradiation lung shielding and dose optimization using asymmetric MLC apertures.

Authors:  Shahbaz Ahmed; Derek Brown; Saad B S Ahmed; Muhammad B Kakakhel; Wazir Muhammad; Amjad Hussain
Journal:  J Appl Clin Med Phys       Date:  2016-03-08       Impact factor: 2.102

7.  Novel rotatable tabletop for total-body irradiation using a linac-based VMAT technique.

Authors:  Christoph Losert; Roel Shpani; Robert Kießling; Philipp Freislederer; Minglun Li; Franziska Walter; Maximilian Niyazi; Michael Reiner; Claus Belka; Stefanie Corradini
Journal:  Radiat Oncol       Date:  2019-12-30       Impact factor: 3.481

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.