Literature DB >> 21448714

Performance of different continuous positive airway pressure helmets equipped with safety valves during failure of fresh gas supply.

Manuela Milan1, Alberto Zanella, Stefano Isgrò, Salua Abd El Aziz El Sayed Deab, Federico Magni, Antonio Pesenti, Nicolò Patroniti.   

Abstract

PURPOSE: We assessed the performance of different continuous positive airway pressure (CPAP) helmets equipped with a safety valve during discontinuation of fresh gas flow.
METHODS: This was a physiological study of five healthy volunteers. We delivered CPAP (fresh gas flow 60 l/min, FiO(2) 60%, PEEP 5 cmH(2)O) with three different helmets in a random sequence: 4Vent (Rüsch), HelmHAR-cp (Harol) and CaStar CP210 (StarMed). For each helmet we randomly applied, for up to 4 min, three disconnections of fresh gas flow: helmet inlet (D(inlet)), flowmeter (D(flowmeter)) and gas source (D(source)). We continuously recorded from a nostril: end-tidal CO(2) (PetCO(2)), inspiratory CO(2) (PiCO(2)), fraction of inspired oxygen (FiO(2)) and respiratory rate (RR).
RESULTS: During every disconnection we observed an increase in PiCO(2) and PetCO(2) with a drop in FiO(2), while RR did not change. FiO(2) decreased more quickly in the CaStar, equipped with the largest safety valve, during D(source) and D(flowmeter), while FiO(2) decreased more quickly during D(inlet) in CaStar and in 4Vent. PiCO(2) resulted in a lower increase in CaStar during D(source) and D(flowmeter) compared to 4Vent. PetCO(2) in CaStar increases more slowly compared to 4Vent during D(source) and more slowly compared to the other two helmets during D(flowmeter). During D(inlet) similar degrees of CO(2) rebreathing and PetCO(2) were recorded among all the helmets.
CONCLUSIONS: To minimize CO(2) rebreathing during disconnection of the fresh gas supply while performing helmet CPAP, it is desirable to utilize large helmets with a large anti-suffocation valve. Monitoring and alarm systems should be employed for safe application of helmet CPAP. © Copyright jointly held by Springer and ESICM 2011

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21448714     DOI: 10.1007/s00134-011-2207-3

Source DB:  PubMed          Journal:  Intensive Care Med        ISSN: 0342-4642            Impact factor:   17.440


  11 in total

Review 1.  Noninvasive ventilation.

Authors:  S Mehta; N S Hill
Journal:  Am J Respir Crit Care Med       Date:  2001-02       Impact factor: 21.405

Review 2.  International Consensus Conferences in Intensive Care Medicine: noninvasive positive pressure ventilation in acute Respiratory failure.

Authors: 
Journal:  Am J Respir Crit Care Med       Date:  2001-01       Impact factor: 21.405

3.  Danger of helmet continuous positive airway pressure during failure of fresh gas source supply.

Authors:  Nicolò Patroniti; Maurizio Saini; Alberto Zanella; Stefano Isgrò; Antonio Pesenti
Journal:  Intensive Care Med       Date:  2006-11-18       Impact factor: 17.440

4.  Home versus intensive care pressure support devices. Experimental and clinical comparison.

Authors:  F Lofaso; L Brochard; T Hang; H Lorino; A Harf; D Isabey
Journal:  Am J Respir Crit Care Med       Date:  1996-05       Impact factor: 21.405

5.  Continuous positive airway pressure delivered with a "helmet": effects on carbon dioxide rebreathing.

Authors:  Paolo Taccone; Dean Hess; Pietro Caironi; Luca M Bigatello
Journal:  Crit Care Med       Date:  2004-10       Impact factor: 7.598

6.  Noninvasive continuous positive airway pressure ventilation using a new helmet interface: a case-control prospective pilot study.

Authors:  Jean-Marie Tonnelier; Gwenaël Prat; Emmanuel Nowak; David Goetghebeur; Anne Renault; Jean Michel Boles; Erwan L'her
Journal:  Intensive Care Med       Date:  2003-11       Impact factor: 17.440

7.  Head helmet versus face mask for non-invasive continuous positive airway pressure: a physiological study.

Authors:  Nicolò Patroniti; Giuseppe Foti; Annamaria Manfio; Anna Coppo; Giacomo Bellani; Antonio Pesenti
Journal:  Intensive Care Med       Date:  2003-08-28       Impact factor: 17.440

8.  Carbon dioxide rebreathing during non-invasive ventilation delivered by helmet: a bench study.

Authors:  Francesco Mojoli; Giorgio A Iotti; Maddalena Gerletti; Carlo Lucarini; Antonio Braschi
Journal:  Intensive Care Med       Date:  2008-04-18       Impact factor: 17.440

Review 9.  The use of helmets to deliver non-invasive continuous positive airway pressure in hypoxemic acute respiratory failure.

Authors:  G Bellani; N Patroniti; M Greco; G Foti; A Pesenti
Journal:  Minerva Anestesiol       Date:  2008-07-18       Impact factor: 3.051

Review 10.  Clinical review: Noninvasive ventilation in the clinical setting--experience from the past 10 years.

Authors:  Massimo Antonelli; Mariano Alberto Pennisi; Luca Montini
Journal:  Crit Care       Date:  2004-08-25       Impact factor: 9.097

View more
  5 in total

Review 1.  Year in review in Intensive Care Medicine 2011: III. ARDS and ECMO, weaning, mechanical ventilation, noninvasive ventilation, pediatrics and miscellanea.

Authors:  Massimo Antonelli; Marc Bonten; Jean Chastre; Giuseppe Citerio; Giorgio Conti; J Randall Curtis; Daniel De Backer; Goran Hedenstierna; Michael Joannidis; Duncan Macrae; Jordi Mancebo; Salvatore M Maggiore; Alexandre Mebazaa; Jean-Charles Preiser; Patricia Rocco; Jean-François Timsit; Jan Wernerman; Haibo Zhang
Journal:  Intensive Care Med       Date:  2012-02-14       Impact factor: 17.440

Review 2.  The use of head helmets to deliver noninvasive ventilatory support: a comprehensive review of technical aspects and clinical findings.

Authors:  Andrea Coppadoro; Elisabetta Zago; Fabio Pavan; Giuseppe Foti; Giacomo Bellani
Journal:  Crit Care       Date:  2021-09-08       Impact factor: 9.097

Review 3.  Clinical review: Helmet and non-invasive mechanical ventilation in critically ill patients.

Authors:  Antonio M Esquinas Rodriguez; Peter J Papadakos; Michele Carron; Roberto Cosentini; Davide Chiumello
Journal:  Crit Care       Date:  2013-04-25       Impact factor: 9.097

4.  Use of a helmet for oxygen therapy in critically ill patients: a systematic review and meta-analysis.

Authors:  Tao Wang; Hongzhen Yin; Qiancheng Xu; Xiaogan Jiang; Tao Yu
Journal:  J Int Med Res       Date:  2020-02       Impact factor: 1.671

5.  Helmet CPAP to Treat Acute Hypoxemic Respiratory Failure in Patients with COVID-19: A Management Strategy Proposal.

Authors:  Dejan Radovanovic; Maurizio Rizzi; Stefano Pini; Marina Saad; Davide Alberto Chiumello; Pierachille Santus
Journal:  J Clin Med       Date:  2020-04-22       Impact factor: 4.241

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.