Comparison of 1H MR spectroscopy, 3-point DIXON, and multi-echo gradient echo for measuring hepatic fat fraction.

Kinya Ishizaka1, Noriko Oyama, Suzuko Mito, Hiroyuki Sugimori, Mitsuhiro Nakanishi, Tomoyuki Okuaki, Hiroki Shirato, Satoshi Terae.   

Abstract

PURPOSE: We evaluated and compared the reproducibility, diagnostic accuracy, and sequence dependency of the fat fraction (FF) determined by 3-point Dixon (DIXON) and multi-echo gradient-echo (MEGE) methods to those of the FF determined by magnetic resonance spectroscopy (MRS).
METHODS: Our study included 98 volunteers, ten of whom underwent scanning twice to evaluate sequence reproducibility. We compared the FFs determined by the DIXON and MEGE methods to that by MRS as the gold standard, calculated sensitivity and specificity for each image analysis method at a threshold value of 6.25% of spectroscopic value, and used Pearson's correlation coefficient and Bland-Altman analysis to compare agreement among the repeated measurements and FF values with the 3 methods, in 98 volunteers.
RESULTS: There was no significant difference in repeated scans in any sequence with Wilcoxon's t-test. Each correlation coefficient (r) exceeded 0.930 for the repeated measurements of all 3 sequences. Sensitivity of DIXON was 82% and specificity, 96%; sensitivity of MEGE was 70% and specificity, 99%. The FFs determined by DIXON and MEGE correlated well with that by MRS (r=0.920) but showed significant difference (paired t-test, P<0.001). The mean difference between the FF determined by DIXON and that by MEGE were 0.93 and -1.16, respectively. The slope of the regression lines as determined by DIXON was -0.655 (P<0.001) and that by MEGE was -0.527 (P<0.001). When the FF by MRS was less than 6.25%, the FF values by DIXON and MEGE were significantly higher; when the spectroscopic value was greater than 6.25%, their values were significantly lower.
CONCLUSION: We demonstrated the high reproducibility of each FF measurement using MRS, DIXON, and MEGE. Compared to MRS, both DIXON and MEGE showed high sensitivity and specificity for determining FF. The FFs by DIXON and MEGE showed sequence dependency because DIXON had proportional and additional errors, and MEGE had a proportional error.

Entities:  

Mesh:

Year:  2011        PMID: 21441727     DOI: 10.2463/mrms.10.41

Source DB:  PubMed          Journal:  Magn Reson Med Sci        ISSN: 1347-3182            Impact factor:   2.471


  5 in total

1.  Liver acquisition with volume acceleration flex on 70-cm wide-bore and 60-cm conventional-bore 3.0-T MRI.

Authors:  Shigeyoshi Saito; Keiko Tanaka; Takashi Hashido
Journal:  Radiol Phys Technol       Date:  2016-01-06

2.  Quantification of early fatty infiltration of the rotator cuff muscles: comparison of multi-echo Dixon with single-voxel MR spectroscopy.

Authors:  Christoph A Agten; Andrea B Rosskopf; Christian Gerber; Christian W A Pfirrmann
Journal:  Eur Radiol       Date:  2015-12-17       Impact factor: 5.315

3.  Heterogeneous living donor hepatic fat distribution on MRI chemical shift imaging.

Authors:  YoungRok Choi; Jeong Min Lee; Nam-Joon Yi; Hyeyoung Kim; Min-Su Park; Geun Hong; Tae Yoo; Suk-Won Suh; Hae Won Lee; Kwang-Woong Lee; Kyung-Suk Suh
Journal:  Ann Surg Treat Res       Date:  2015-06-11       Impact factor: 1.859

4.  Quantitative magnetic resonance imaging evaluation of hepatic fat content with iron deposition: will it be disturbed?

Authors:  Qian Wang; Feng Ye; Peiqing Ma; Feng Chen; Yiqun Che; Xinming Zhao; Li Yang
Journal:  J Int Med Res       Date:  2019-03-12       Impact factor: 1.671

5.  Dynamics of hepatic steatosis resolution and changes in gut microbiome with weight loss in nonalcoholic fatty liver disease.

Authors:  Jaideep Behari; Lisa Graham; Renwei Wang; Claudiu Schirda; Amir A Borhani; Barbara A Methé; Kelvin Li; Alison Morris; Hung N Luu; Sabrina Palmieri; Jian-Min Yuan
Journal:  Obes Sci Pract       Date:  2021-01-04