Literature DB >> 21434958

Stability of hematologic analytes in monkey, rabbit, rat, and mouse blood stored at 4°C in EDTA using the ADVIA 120 hematology analyzer.

Mehrdad Ameri1, Henry A Schnaars, John R Sibley, David J Honor.   

Abstract

BACKGROUND: The time from sampling to analysis can be delayed when blood samples are shipped to distant reference laboratories or when analysis cannot be readily performed.
OBJECTIVE: The objective of this study was to evaluate the stability of hematologic analytes in blood samples from monkeys, rabbits, rats, and mice when samples were stored for up to 72 hours at 4°C.
METHODS: Blood samples from 30 monkeys, 15 rabbits, 20 rats, and 30 mice were collected into EDTA-containing tubes and were initially analyzed within 1 hour of collection using the ADVIA 120 analyzer. The samples were then stored at 4°C and reanalyzed at 24, 48, and 72 hours after collection.
RESULTS: Significant (P<.0003) changes in hematologic analytes and calculations included increased HCT and MCV and decreased MCHC and cell hemoglobin concentration mean (CHCM) at 72 hours and increased MPV at 24 hours in monkeys; increased MCV at 72 hours and MPV at 48 hours and decreased monocyte count at 24 hours in rabbits; increased MCV and decreased MCHC, CHCM, and monocyte count at 24 hours in rats; increased MCV, red cell distribution width, and MPV and decreased MCHC, CHCM, and monocyte count at 24 hours in mice.
CONCLUSIONS: Although most of the changes in the hematologic analytes in blood from monkeys, rabbits, rats, and mice when samples were stored at 4°C were analytically acceptable and clinically negligible, the best practice in measuring hematologic analytes in these animals is timely processing of blood samples, preferably within 1 hour after collection. ©2011 American Society for Veterinary Clinical Pathology.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21434958     DOI: 10.1111/j.1939-165X.2011.00304.x

Source DB:  PubMed          Journal:  Vet Clin Pathol        ISSN: 0275-6382            Impact factor:   1.180


  3 in total

Review 1.  Practical murine hematopathology: a comparative review and implications for research.

Authors:  Karyn E O'Connell; Amy M Mikkola; Aaron M Stepanek; Andyna Vernet; Christopher D Hall; Chia C Sun; Eda Yildirim; John F Staropoli; Jeannie T Lee; Diane E Brown
Journal:  Comp Med       Date:  2015-04       Impact factor: 0.982

2.  Artifactual changes in Sprague-Dawley rat hematologic parameters after storage of samples at 3 °C and 21 °C.

Authors:  Michelle C Cora; Debra King; Laura J Betz; Ralph Wilson; Greg S Travlos
Journal:  J Am Assoc Lab Anim Sci       Date:  2012       Impact factor: 1.232

3.  Comparison of Two Methods for the Determination of the Effects of Ionizing Radiation on Blood Cell Counts in Mice.

Authors:  Ana L Romero-Weaver; Ann R Kennedy
Journal:  Int J Biomed Sci       Date:  2012-03-15
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.