Literature DB >> 21420591

Femtosecond laser flap parameters and visual outcomes in laser in situ keratomileusis.

Ammar Issa1, Usama Al Hassany.   

Abstract

PURPOSE: To evaluate the flap-producing characteristics of the VisuMax femtosecond laser system and their relationship to visual and refractive outcomes after laser in situ keratomileusis (LASIK).
SETTING: Dr. Sulaiman Al Habib Medical Center, Dubai Healthcare City, Dubai, United Arab Emirates.
DESIGN: Case series.
METHODS: A review of LASIK cases from December 2008 to December 2009 was performed. Parameters recorded included the achieved flap thickness measured with intraoperative ultrasonic pachymetry, laser energy, side-cut angle, visual outcomes including uncorrected (UDVA) and corrected (CDVA) distance visual acuities and manifest refraction, higher-order aberrations using Hartman-Shack wavefront analysis, and complications.
RESULTS: The study comprised 98 patients (195 eyes). The femtosecond laser energy for flap preparation was 160 to 200 nJ. The mean achieved flap thickness was 91.35 μm ± 12.26 (SD) (attempted 80 μm), 98.5 ± 14.39 μm (attempted 90 μm), 109.94 ± 13.43 μm (attempted 100 μm), 112.7 ± 7.45 μm (attempted 110 μm), and 123.82 ± 9.64 μm (attempted 120 μm). Postoperatively, 87% of eyes were within ± 0.25 diopter of the intended spherical equivalent, 75% had a UDVA of 20/16 (-0.1 logMAR) or better, and 42% gained 1 or more lines of visual acuity; no eye lost CDVA. No statistically significant relationships were found between energy, side-cut angle, achieved flap thickness, and refractive and visual outcomes. Complications were mild and did not affect final visual outcomes.
CONCLUSION: The femtosecond laser yielded precise flap dimensions with a narrow standard deviation and a high level of safety. FINANCIAL DISCLOSURE: Neither author has a financial or proprietary interest in any material or method mentioned.
Copyright © 2011 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21420591     DOI: 10.1016/j.jcrs.2010.10.049

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  7 in total

1.  Laser-assisted in-situ keratomileusis (LASIK) with a mechanical microkeratome compared to LASIK with a femtosecond laser for LASIK in adults with myopia or myopic astigmatism.

Authors:  Nicolás Kahuam-López; Alejandro Navas; Carlos Castillo-Salgado; Enrique O Graue-Hernandez; Aida Jimenez-Corona; Antonio Ibarra
Journal:  Cochrane Database Syst Rev       Date:  2020-04-07

2.  Comparison of visual performance recovery after thin-flap LASIK with 4 femtosecond lasers.

Authors:  Jing Zhang; Yue-Hua Zhou; Yan Zheng; Qian Liu
Journal:  Int J Ophthalmol       Date:  2017-10-18       Impact factor: 1.779

3.  Comparative study of visual acuity and aberrations after intralase femtosecond LASIK: small corneal flap versus big corneal flap.

Authors:  Ya-Li Zhang; Lei Liu; Chang-Xia Cui; Ming Hu; Zhao-Na Li; Li-Jun Cao; Xiu-Hua Jing; Guo-Ying Mu
Journal:  Int J Ophthalmol       Date:  2013-10-18       Impact factor: 1.779

4.  Comparison of the femtosecond laser and mechanical microkeratome for flap cutting in LASIK.

Authors:  Li-Kun Xia; Jie Yu; Guang-Rui Chai; Dang Wang; Yang Li
Journal:  Int J Ophthalmol       Date:  2015-08-18       Impact factor: 1.779

5.  Using femtosecond laser to create customized corneal flaps for patients with low and moderate refractive error differing in corneal thickness.

Authors:  Chi Zhang; Jingbin Che; Jianhong Yu; Linli Yu; Dan Yu; Gangping Zhao
Journal:  PLoS One       Date:  2015-03-25       Impact factor: 3.240

6.  Visumax femtolasik versus Moria M2 microkeratome in mild to moderate myopia: efficacy, safety, predictability, aberrometric changes and flap thickness predictability.

Authors:  Magda A Torky; Yousif A Al Zafiri; Abeer M Khattab; Rania K Farag; Eman A Awad
Journal:  BMC Ophthalmol       Date:  2017-07-17       Impact factor: 2.209

Review 7.  Small incision lenticule extraction (SMILE) history, fundamentals of a new refractive surgery technique and clinical outcomes.

Authors:  Dan Z Reinstein; Timothy J Archer; Marine Gobbe
Journal:  Eye Vis (Lond)       Date:  2014-10-16