Parameter signature of a reel problem.

Frazier E Fyke1, Sandra S McCearley.   

Abstract

Clinics involved in follow-up of cardiac implantable electronic devices, especially implantable cardiac defibrillators (ICDs), increasing rely upon remote monitoring. This case describes the parameter signature characteristic of an uncommon but well-described syndrome. Analysis of the nature and timing of the sequence of abnormalities contained in the archived device data, all of which is available for review via remote monitoring, reveals the cause for failure of this primary prevention ICD system. ©2011, The Authors. Journal compilation ©2011 Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2011        PMID: 21410730     DOI: 10.1111/j.1540-8159.2011.03055.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  1 in total

1.  Delayed presentation of pacemaker Reel Syndrome and lead damage: Coincidence or consequence?

Authors:  Safae Hilal; Fatima Chikhi; Fadoum Hassan; Ibtissam Fellat; Mohammed Cherti
Journal:  Indian Pacing Electrophysiol J       Date:  2022-07-09