| Literature DB >> 21394848 |
Isabela Costa César1, Ricardo Martins Duarte Byrro, Fabiana Fernandes de Santana e Silva Cardoso, Iram Moreira Mundim, Leonardo de Souza Teixeira, Weidson Carlo de Sousa, Sandro Antônio Gomes, Karini Bruno Bellorio, Juliana Machado Brêtas, Gerson Antônio Pianetti.
Abstract
A liquid chromatography-electrospray ionization tandem mass spectrometry (HPLC-ESI-MS/MS) method for the quantitation of triamcinolone in human plasma after nasal spray application was developed and validated. Betamethasone was used as internal standard (IS). The analytes were extracted by a liquid-liquid procedure and separated on a Zorbax Eclipse XDB C(18) column with a mobile phase composed of 2 mM aqueous ammonium acetate pH 3.2 and acetonitrile (55:45). Selected reaction monitoring was performed using the transitions m/z 435 → 415 and m/z 393 → 373 to quantify triamcinolone acetonide and betamethasone, respectively. Calibration curve was constructed over the range of 20-2000 pg/ml for triamcinolone acetonide. The lower limit of quantitation was 20 pg/ml. The mean RSD values were 4.6% and 5.7% for the intra-run and inter-run precision, respectively. The mean accuracy value was 98.5% and a recovery rate corresponding to 97.5% was achieved. No matrix effect was detected in the samples. The validated method was successfully applied to determine the plasma concentrations of triamcinolone acetonide in healthy volunteers, in a pharmacokinetic study with nasal spray formulation.Entities:
Mesh:
Substances:
Year: 2011 PMID: 21394848 DOI: 10.1002/jms.1896
Source DB: PubMed Journal: J Mass Spectrom ISSN: 1076-5174 Impact factor: 1.982