Evaluation of four commercial biuret reagent kits of serum total protein by the American Association for Clinical Chemistry reference measurement procedure.

Meilin He1, Jie Zhang.   

Abstract

BACKGROUND: In China, the traceability of clinical chemistry methods is still immature. Therefore, it is necessary to establish a reference measurement procedure and evaluate commercial reagent kits using such established procedures.
METHODS: We reproduced the reference measurement procedure for serum total protein, as recommended by the American Association for Clinical Chemistry (AACC). We evaluated the performance by Clinical Laboratory Standard Institute (CLSI) guidelines EP15-A and EP6-A. Subsequently, four commercial reagent kits were evaluated by the reproduced reference procedure following CLSI guideline EP9-A2.
RESULTS: The performance of the reproduced reference procedure was as follows: CVs ranged from 0.47% to 0.85% at medical decision levels (X(c)) of 45 g/L, 60 g/L and 80 g/L. Linearity was Y=1.0022X-0.2121 (r=0.9999), and recovery ranged from 100.2% to 102.4%. The External Quality Assessment Scheme for Reference Laboratories in Laboratory Medicine (RELA) was applied, and the result was within the limit of equivalence. The linear relationships of four commercial reagent kits, Merit Choice, KHB, Leadman, and Olympus, were, respectively: Y=0.9922X+0.5776 (r=0.9961); Y=0.9936X+0.4316 (r=0.9992); Y=0.9949X+0.9129 (r=0.9987) and Y=0.9923X+0.8876 (r=0.9989). KHB showed slight negative bias, and the mean±SD was -0.03±0.60 g/L. Merit Choice, Leadman, and Olympus all showed positive bias, and the mean±SDs were 0.02±0.63 g/L, 0.55±0.77 g/L and 0.34±0.71 g/L, respectively.
CONCLUSIONS: The correlation and bias of four commercial reagent kits for serum total protein were found to be acceptable. Thus, these reagent kits can be used reliably in China.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21391866     DOI: 10.1515/CCLM.2011.153

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  1 in total

1.  Distribution of serum total protein in elderly Chinese.

Authors:  Chang-Rong Tian; Li Qian; Xiao-Zhu Shen; Jia-Jing Li; Jiang-Tao Wen
Journal:  PLoS One       Date:  2014-06-26       Impact factor: 3.240

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.