Early and midterm results of totally endoscopic coronary artery bypass grafting on the beating heart.

Changqing Gao1, Ming Yang, Yang Wu, Gang Wang, Cangsong Xiao, Yue Zhao, Jiali Wang.   

Abstract

OBJECTIVE: Despite the early introduction of totally endoscopic coronary artery bypass on the beating heart, only a limited number of cases have been performed. The limiting factor has been the concern about safety and graft patency of the anastomosis. This study describes our experience with totally endoscopic coronary artery bypass on the beating heart with robotic assistance and its early and midterm results.
METHODS: In 365 cases of robotic cardiac operations, 162 patients underwent robotic coronary artery bypass grafting on the beating heart, of whom 60 patients (46 male, 14 female) underwent totally endoscopic coronary artery bypass on the beating heart. The patients' mean age was 56.97 ± 9.7 years (33-77 years). Left internal thoracic artery to left anterior descending anastomosis was performed using the U-Clip device.
RESULTS: We completed 58 totally endoscopic coronary artery bypass procedures, in which 16 patients received hybrid procedures. Two patients had conversions to a minithoracotomy. The average left internal thoracic artery harvesting and anastomosis times were 31.3 ± 10.5 (18∼55) minutes and 11.3 ± 4.7 (5∼21) minutes, respectively. The mean operating room and operation times were 336.1 ± 58.5 (210∼580) minutes and 264.8 ± 65.6 (150∼420) minutes, respectively. The drainage was 164.9 ± 83.2 (70∼450) mL. Before discharge, 50 patients underwent angiography and 8 patients underwent computed tomography angiography, and the study showed that graft patency was 100%. Unexpectedly, the left internal thoracic artery graft developed a collateral branch in 2 patients. After discharge, all patients were followed up by computed tomography angiography. The average follow-up time was 12.67 ± 9.43 (1-40) months. One patient had gastric bleeding after surgery.
CONCLUSIONS: Totally endoscopic coronary artery bypass on the beating heart is a safe procedure in selected patients and produces excellent early and midterm patency of anastomosis.
Copyright © 2011 The American Association for Thoracic Surgery. Published by Mosby, Inc. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21388642     DOI: 10.1016/j.jtcvs.2011.01.051

Source DB:  PubMed          Journal:  J Thorac Cardiovasc Surg        ISSN: 0022-5223            Impact factor:   5.209


  9 in total

1.  Robotic totally endoscopic coronary artery bypass for isolated ostial stenosis of the left coronary artery.

Authors:  Chia-Cheng Kuo; Chung-Hsi Hsing; Chun-I Chen; Chen-Hui Lee; Bor-Chih Cheng
Journal:  J Thorac Dis       Date:  2017-12       Impact factor: 2.895

Review 2.  Videoscope-assisted cardiac surgery.

Authors:  Kuan-Ming Chiu; Robert Jeen-Chen Chen
Journal:  J Thorac Dis       Date:  2014-01       Impact factor: 2.895

3.  Initial experience with internal mammary artery harvesting with the da Vinci Surgical System for minimally invasive direct coronary artery bypass.

Authors:  Tomoyuki Fujita; Hiroki Hata; Yusuke Shimahara; Shunsuke Sato; Junjiro Kobayashi
Journal:  Surg Today       Date:  2014-01-15       Impact factor: 2.549

Review 4.  Is totally endoscopic coronary artery bypass safe, feasible and effective?

Authors:  Metesh Nalin Acharya; Hutan Ashrafian; Thanos Athanasiou; Roberto Casula
Journal:  Interact Cardiovasc Thorac Surg       Date:  2012-09-12

5.  Systematic review of robotic-assisted, totally endoscopic coronary artery bypass grafting.

Authors:  Michael Seco; J James B Edelman; Tristan D Yan; Michael K Wilson; Paul G Bannon; Michael P Vallely
Journal:  Ann Cardiothorac Surg       Date:  2013-07

Review 6.  Anesthetic issues for robotic cardiac surgery.

Authors:  Wendy K Bernstein; Andrew Walker
Journal:  Ann Card Anaesth       Date:  2015 Jan-Mar

7.  Minimally Invasive Cardiac Surgery in China: Multi-Center Experience.

Authors:  Kun Hua; Yang Zhao; Ran Dong; Taoshuai Liu
Journal:  Med Sci Monit       Date:  2018-01-22

8.  Robotic totally endoscopic coronary artery bypass grafting: It's now or never!

Authors:  Husam H Balkhy
Journal:  JTCVS Tech       Date:  2021-04-10

9.  Eleven-year outcomes of U-clips in totally robotic coronary artery bypass grafting versus standard hand-sewn running suture in robotic-assisted coronary artery bypass grafting.

Authors:  Nan Cheng; Huajun Zhang; Ming Yang; Guopeng Liu; Yi Guo; Wenbin Kang; Changqing Gao; Rong Wang
Journal:  Interact Cardiovasc Thorac Surg       Date:  2021-06-28
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.