Literature DB >> 21388287

Pedicle screw insertion angle and pullout strength: comparison of 2 proposed strategies.

Serkan Inceoğlu1, William H Montgomery, Selvon St Clair, Robert F McLain.   

Abstract

OBJECT: Minimally invasive pedicle screws inserted vertically (that is, dorsoventrally) through the pedicle, as opposed to the more common coaxial technique, offer potential advantages by minimizing soft-tissue stripping during screw placement. The screws are designed for insertion through a medial starting point with vertical trajectory through the pedicle and into the vertebral body. As such, no lateral dissection beyond the insertion point is necessary. However, the effects of this insertion technique on the screw biomechanical performance over a short- and long-term are unknown. The authors investigated the pullout strength and stiffness of these screws, with or without fatigue cycling, compared with comparably sized, traditional screws placed by coaxial technique.
METHODS: Twenty-one lumbar vertebrae (L-3, L-4, and L-5) were tested. Each pedicle of each vertebra was instrumented with either a traditional, coaxial pedicle screw (Group A), placed through a standard starting point, or a vertically oriented, alternative-design screw (Group B), with a medial starting point and vertical trajectory. The specimens were divided into 2 groups for testing. One group was tested for direct pullout (10 specimens) while the other was subject to pullout after tangential (toggle) cyclic loading (11 specimens). The screws were cycled in displacement control (± 5 mm producing ~ 4-Nm moment) at a rate of 3 Hz for 5000 cycles. Pullout tests were performed at a rate of 1 mm/minute.
RESULTS: Two-way ANOVA showed that Group B screws with a medial starting point (2541 ± 1090 N for cycled vs 2135 ± 1323 N for noncycled) had significantly higher pullout loads than Group A screws with a standard entry point (1585 ± 766 N for cycled vs 1417 ± 812 N noncycled) (p = 0.001). There was no significant effect of cycling or screw insertion type on pullout stiffness. Tangential stiffness of the Group B screws was significantly less than that of the Group A screws (p = 0.001). The stiffness of both screws in the toe region was significantly affected by cycling (p = 0.001).
CONCLUSIONS: The use of Group B screws inserted through a medial starting point showed greater pullout load than a Group A screw inserted through a standard starting point. The greater pullout strength in Group B screws may be due to screw thread design and increased cortical bone purchase at the medial starting point. Nevertheless, anatomical considerations of the medial starting point, that is, pedicle or lateral vertebral body cortex breach, may limit its application. The medial starting point of the Group B screw was frequently in the facet at the L-3 and L-4 pedicle entry points, which may have clinical importance.

Entities:  

Mesh:

Year:  2011        PMID: 21388287     DOI: 10.3171/2010.11.SPINE09886

Source DB:  PubMed          Journal:  J Neurosurg Spine        ISSN: 1547-5646


  20 in total

1.  Morphometric measurement of the lumbosacral spine for minimally invasive cortical bone trajectory implant using computed tomography.

Authors:  Hua Zhang; Remi Musibau Ajiboye; Arya Nick Shamie; Qionghua Wu; Qixin Chen; Weishan Chen
Journal:  Eur Spine J       Date:  2015-09-05       Impact factor: 3.134

2.  Cortical and Standard Trajectory Pedicle Screw Fixation Techniques in Stabilizing Multisegment Lumbar Spine with Low Grade Spondylolisthesis.

Authors:  Wayne K Cheng; Serkan İnceoğlu
Journal:  Int J Spine Surg       Date:  2015-08-31

3.  Quantitative dual-energy CT for phantomless evaluation of cancellous bone mineral density of the vertebral pedicle: correlation with pedicle screw pull-out strength.

Authors:  Julian L Wichmann; Christian Booz; Stefan Wesarg; Ralf W Bauer; J Matthias Kerl; Sebastian Fischer; Thomas Lehnert; Thomas J Vogl; M Fawad Khan; Konstantinos Kafchitsas
Journal:  Eur Radiol       Date:  2014-12-07       Impact factor: 5.315

4.  First clinical results of minimally invasive vector lumbar interbody fusion (MIS-VLIF) in spondylodiscitis and concomitant osteoporosis: a technical note.

Authors:  Bernhard Rieger; Hongzhen Jiang; Daniel Ruess; Clemens Reinshagen; Marek Molcanyi; Jozef Zivcak; Huaiyu Tong; Gabriele Schackert
Journal:  Eur Spine J       Date:  2016-12-27       Impact factor: 3.134

5.  Effect of screw position on load transfer in lumbar pedicle screws: a non-idealized finite element analysis.

Authors:  Anna G U S Newcomb; Seungwon Baek; Brian P Kelly; Neil R Crawford
Journal:  Comput Methods Biomech Biomed Engin       Date:  2016-07-25       Impact factor: 1.763

6.  Comparing the Biomechanical Stability of Cortical Screw Trajectory Versus Standard Pedicle Screw Trajectory for Short- and Long-Segment Posterior Fixation in 3-Column Thoracic Spinal Injury.

Authors:  Amey R Savardekar; Nestor G Rodriguez-Martinez; Anna G U S Newcomb; Phillip M Reyes; Hector Soriano-Baron; Steve W Chang; Brian P Kelly; Neil R Crawford
Journal:  Int J Spine Surg       Date:  2019-06-30

7.  Pullout force of minimally invasive surgical and open pedicle screws-a biomechanical cadaveric study.

Authors:  Phoebe G M Matthews; Joseph Cadman; Janos Tomka; Danè Dabirrahmani; Richard Appleyard; Andrew Kam
Journal:  J Spine Surg       Date:  2020-03

8.  Comparison of three different screw trajectories in osteoporotic vertebrae: a biomechanical investigation.

Authors:  J-S Jarvers; S Schleifenbaum; C Pfeifle; C Oefner; M Edel; N von der Höh; C-E Heyde
Journal:  BMC Musculoskelet Disord       Date:  2021-05-05       Impact factor: 2.362

9.  Cortical bone trajectory screws for the middle-upper thorax: An anatomico-radiological study.

Authors:  Sun-Ren Sheng; Jiao-Xiang Chen; Wei Chen; En-Xing Xue; Xiang-Yang Wang; Qing-An Zhu
Journal:  Medicine (Baltimore)       Date:  2016-08       Impact factor: 1.889

10.  Midline lumbar fusion with cortical bone trajectory screw.

Authors:  Masaki Mizuno; Keita Kuraishi; Yasuyuki Umeda; Takanori Sano; Masanori Tsuji; Hidenori Suzuki
Journal:  Neurol Med Chir (Tokyo)       Date:  2014-08-29       Impact factor: 1.742

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.