Yusuke Iwasaki1, Keisuke Mochizuki, Yuki Nakano, Natsumi Maruya, Masato Goto, Yosuke Maruyama, Rie Ito, Koichi Saito, Hiroyuki Nakazawa.
Abstract
Reactive oxygen species (ROS) and reactive nitrogen species (RNS) are well-known and important contributors to oxidative and nitrosative stress in several diseases. Hydroxylated phenylalanine and nitrated tyrosine products appear to be particularly susceptible targets of oxidative and nitrosative stress. We compared fluorescence reagents for their potential use in the analysis of hydroxylated phenylalanine and nitrated tyrosine products with a high-sensitivity and high-specificity HPLC-UV-FL technique. The analytes were extracted from serum via solid-phase extraction on Waters Oasis MCX cartridges. Chromatographic separation was achieved on an ODS column (Capcell Pak MG II; 150 × 2.0 mm) using a gradient mobile phase consisting of 20 mm sodium phosphate buffer (adjusted to pH 3.0) and acetonitrile. The method quantification limit for 4-nitrophenylalanine, m-tyrosine, and 3-nitrotyrosine was 0.1 μm. The relative standard deviation of the precision and accuracy was acceptable at the spiked concentration of 0.1 μm for 4-nitrophenylalanine, m-tyrosine and 3-nitrotyrosine. The method could be used for the in vitro analysis of serum samples.Entities:
Mesh:
Substances:
Year: 2011 PMID: 21387354 DOI: 10.1002/bmc.1623