Dangers of bone graft substitutes: lessons from using GeneX.

Samira Saadoun1, Claire Macdonald, B Anthony Bell, Marios C Papadopoulos.   

Abstract

BACKGROUND: Bone graft substitutes are widely used in spinal surgery. Here, serious complications associated with the bone graft substitute GeneX are presented. GeneX contains tri-calcium phosphate and calcium sulphate.
METHODS: GeneX was used in three patients who had spinal decompression and fusion. Mice were also injected with GeneX, demineralised bone matrix (DBX) or saline subcutaneously. After 24 h the extent of tissue damage and inflammation in tissue sections was quantified. To understand the licensing process for bone graft substitutes, the U.S. Food and Drug Administration (FDA) and the U.K. Medicines and Healthcare products Regulatory Agency (MHRA) websites were accessed.
RESULTS: All patients developed sterile pus in soft tissues adjacent to the GeneX followed by skin breakdown in two and pharyngeal perforation in one. In mice, GeneX produced moderate or severe skin damage compared with no or mild skin damage after DBX (p<0.05) or saline (p<0.05) injection. GeneX caused more inflammation in mouse dermis (1704±193 leucocytes/mm2, mean ± SE) than DBX (537 ± 266, p<0.01) or saline (136 ± 19, p<0.01). The FDA and MHRA classify bone graft substitutes as medical devices. In contrast with drugs, medical devices do not need to undergo clinical safety tests before obtaining FDA 510(k) clearance for use in patients.
CONCLUSION: GeneX may cause soft tissue inflammation and destruction and should not be placed next to thin walled structures, such as skin or pharynx, because it may erode through these tissues. Bone graft substitutes should undergo mandatory detailed safety testing prior to approval. This could be achieved by reclassifying them as drugs.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21386107     DOI: 10.1136/jnnp.2010.232371

Source DB:  PubMed          Journal:  J Neurol Neurosurg Psychiatry        ISSN: 0022-3050            Impact factor:   10.154


  8 in total

1.  Reply to the Letter to the Editor: Adverse reactions of artificial bone graft substitutes: lessons learned from using tricalcium phosphate geneX ®.

Authors:  Joerg Friesenbichler; Werner Maurer-Ertl; Patrick Sadoghi; Ulrike Pirker-Fruehauf; Koppany Bodo; Andreas Leithner
Journal:  Clin Orthop Relat Res       Date:  2013-12-10       Impact factor: 4.176

2.  Primary cilia are sensors of electrical field stimulation to induce osteogenesis of human adipose-derived stem cells.

Authors:  Shaobo Cai; Josephine C Bodle; Pattie S Mathieu; Alison Amos; Mehdi Hamouda; Susan Bernacki; Greg McCarty; Elizabeth G Loboa
Journal:  FASEB J       Date:  2016-10-19       Impact factor: 5.191

3.  Adverse reactions of artificial bone graft substitutes: lessons learned from using tricalcium phosphate geneX®.

Authors:  Joerg Friesenbichler; Werner Maurer-Ertl; Patrick Sadoghi; Ulrike Pirker-Fruehauf; Koppany Bodo; Andreas Leithner
Journal:  Clin Orthop Relat Res       Date:  2014-03       Impact factor: 4.176

4.  Lytic Complications after Skull Reconstruction Using GeneX®.

Authors:  Jin-Hack Park; Sang-Jun Suh; Yoon-Soo Lee; Jeong-Ho Lee; Kee-Young Ryu; Dong-Gee Kang
Journal:  Korean J Neurotrauma       Date:  2015-10-31

5.  Adverse effect of beta-tricalcium phosphate with zeta potential control in repairing critical defects in rats' calvaria.

Authors:  Daniel Falbo Martins de Souza; Luciana Correa; Daniel Isaac Sendyk; Rafael Augusto Burim; Maria da Graça Naclério-Homem; Maria Cristina Zindel Deboni
Journal:  Rev Bras Ortop       Date:  2016-04-26

6.  Clinical experience with the artificial bone graft substitute Calcibon used following curettage of benign and low-grade malignant bone tumors.

Authors:  Joerg Friesenbichler; Werner Maurer-Ertl; Marko Bergovec; Lukas A Holzer; Kathrin Ogris; Lukas Leitner; Andreas Leithner
Journal:  Sci Rep       Date:  2017-05-11       Impact factor: 4.379

7.  Adverse reactions of artificial bone graft substitutes: lessons learned from using tricalcium phosphate geneX ®.

Authors:  Phillip Anthony Laycock; John J Cooper
Journal:  Clin Orthop Relat Res       Date:  2013-12-10       Impact factor: 4.176

8.  Clinical Trial Assessment Principles of National Class III Medical Devices in China.

Authors:  Yue Min; Jin-Tao Gao; Jing Wu; Bao Zhai; Dan Han; Bin Liu
Journal:  Orthop Surg       Date:  2019-09-06       Impact factor: 2.071

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.