Measurement of exhaled nitric oxide in a general population sample: a comparison of the Medisoft HypAir FE(NO) and Aerocrine NIOX analyzers.

Collin R Brooks1, Shirley-Belle M Brogan, Christine J van Dalen, Phillipa K Lampshire, Julian Crane, Jeroen Douwes.   

Abstract

BACKGROUND AND
OBJECTIVE: Measuring the fraction of nitric oxide in exhaled breath (FE(NO)) is increasingly utilized to assess airway inflammation in asthma. The primary aim of this study was to compare exhaled nitric oxide measurements obtained using two devices from different manufacturers, that is, the recently marketed portable and electrochemical-based Medisoft HypAir FE(NO) and the well-established chemiluminescence-based Aerocrine NIOX analyzer, in an unselected population.
METHODS: FE(NO) measurements were conducted in 106 subjects (86 healthy; 20 asthmatic; 56.6% atopic). Atopy and health status were assessed by skin prick tests and questionnaire, respectively.
RESULTS: The two instruments showed strong correlation over a wide range of FE(NO) measurements (8-261.3 ppb with the HypAir, 5.6-156.8 ppb with the NIOX; r = 0.98; p < .0001). This correlation was observed in the population as a whole, as well as in healthy non-atopics, healthy atopics, and atopic asthmatics when considered separately. The measurements on the HypAir FE(NO) were consistently 1.6 times (95% CI 1.11-2.05) higher than those obtained with the NIOX.
CONCLUSIONS: FE(NO) measurements obtained with the HypAir FE(NO) correlated well with the NIOX, but were approximately 1.6 times higher. Therefore, a conversion factor is required if results are to be compared with the NIOX instrument.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21385111     DOI: 10.3109/02770903.2011.560322

Source DB:  PubMed          Journal:  J Asthma        ISSN: 0277-0903            Impact factor:   2.515


  4 in total

Review 1.  Noninvasive effects measurements for air pollution human studies: methods, analysis, and implications.

Authors:  Jaime Mirowsky; Terry Gordon
Journal:  J Expo Sci Environ Epidemiol       Date:  2015-01-21       Impact factor: 5.563

2.  Measurement of Exhaled Nitric Oxide in 456 Lung Cancer Patients Using a Ringdown FENO Analyzer.

Authors:  Jing Li; Qingyuan Li; Xin Wei; Qing Chen; Meixiu Sun; Yingxin Li
Journal:  Metabolites       Date:  2021-05-31

3.  Fraction of exhaled nitric oxide (FeNO ) norms in healthy Tunisian adults.

Authors:  Sonia Rouatbi; Mohamed Ali Chouchene; Ines Sfaxi; Mohamed Ben Rejeb; Zouhair Tabka; Helmi Ben Saad
Journal:  Biomed Res Int       Date:  2014-06-03       Impact factor: 3.411

4.  Enhanced airway sensory nerve reactivity in non-eosinophilic asthma.

Authors:  Hajar Ali; Collin Brooks; Julian Crane; Richard Beasley; Stephen Holgate; Peter Gibson; Philip Pattemore; Yu-Chieh Tzeng; Thorsten Stanley; Neil Pearce; Jeroen Douwes
Journal:  BMJ Open Respir Res       Date:  2021-11
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.