Literature DB >> 21376404

Pocket related complications in 163 patients receiving anticoagulation or dual antiplatelet therapy: D-Stat Hemostat versus standard of care.

Marc-Alexander Ohlow1, Bernward Lauer, Bjoern Buchter, Matthias Schreiber, J Christoph Geller.   

Abstract

BACKGROUND: Pocket hematoma is a common complication after cardiac rhythm device implantation (CRDI) in patients receiving anticoagulation or dual antiplatelet therapy (DAPT). We sought to assess the safety and effectiveness of D-Stat Flowable Hemostat ([D-Stat]-a mixture of thrombin and collagen) in these patients.
METHODS: We conducted a prospective case control study and enrolled all patients admitted for CRDI receiving anticoagulation or DAPT. Participants received either D-Stat into the pectoral pocket as an adjunct to standard of care versus the standard of care (UC) alone. The primary endpoint was a composite of hematoma needing evacuation, and pocket infection.
RESULTS: We included a total of 163 patients (mean age 72 ± 8 years). 119 (73%) were male. 82 patients (50%) received D-stat and 81 (50%) usual care; 82 patients (50%) were treated with anticoagulation, 62 patients (38%) received DAPT, and 19 patients (12%) both. The primary endpoint occurred in 12/82 patients from the D-Stat group and 3/81 patients of the control group (14.6% versus 3.7%; p=0.03). The use of D-Stat tended for a higher incidence of pocket hematoma requiring evacuation (8.5% versus 2.4%; p=0.12), and an increased rate of pocket infections (6.1% versus 1.2%; p=0.21). Postoperative Immunoglobulin E levels (86 ± 147 IU/ml (D-Stat) versus 120 ± 231 IU/ml (UC)) and increase of C-reactive protein (D-Stat: ∆19 mg/l versus UC: 14 mg/l) were similar in both groups (p=0.26 and p=0.4, respectively).
CONCLUSIONS: Administration of D-Stat does not decrease the frequency of clinically relevant pocket hematomas. The observed rate of pocket infections in the D-Stat group is a matter of concern.
Copyright © 2011 Elsevier Ireland Ltd. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21376404     DOI: 10.1016/j.ijcard.2011.02.042

Source DB:  PubMed          Journal:  Int J Cardiol        ISSN: 0167-5273            Impact factor:   4.164


  9 in total

1.  [Prevention of pocket-related complications following heart rhythm device implantation. D-Stat Hemostat™ versus vacuum drainage].

Authors:  Marc-Alexander Ohlow; Björn Buchter; Michele Brunelli; Bernward Lauer; Matthias Schreiber; J Christoph Geller
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2015-02-03

2.  Pocket related complications following cardiac electronic device implantation in patients receiving anticoagulation and/or dual antiplatelet therapy: prospective evaluation of different preventive strategies.

Authors:  Hassan Awada; J Christoph Geller; Michele Brunelli; Marc-Alexander Ohlow
Journal:  J Interv Card Electrophysiol       Date:  2018-11-21       Impact factor: 1.900

Review 3.  Risk of bleeding associated with interventional musculoskeletal radiology procedures. A comprehensive review of the literature.

Authors:  Gregory B Foremny; Juan Pretell-Mazzini; Jean Jose; Ty K Subhawong
Journal:  Skeletal Radiol       Date:  2014-11-30       Impact factor: 2.199

Review 4.  Arguments to Apply Epinephrine for Pocket Hematoma Reduction. The MAITRE Study.

Authors:  Nikolay Ilov; Nikolay Ilov; Anatoly Nechepurenko; Albert Abdulkadyrov; Damir Paskeev; Elena Damrina; Elena Kulikova; Marina Terent'eva; Dinara Stompel; Dmitry Tarasov
Journal:  J Atr Fibrillation       Date:  2016-06-30

Review 5.  Prevention of Cardiac Implantable Electronic Device Infections: Update and Evaluation of the Potential Role for Capsulectomy or the Antibiotic Pouch.

Authors:  Marye J Gleva; Jeanne E Poole
Journal:  J Atr Fibrillation       Date:  2017-02-28

6.  Prevention of cardiac implantable electronic device infections: guidelines and conventional prophylaxis.

Authors:  Carina Blomstrom-Lundqvist; Bozena Ostrowska
Journal:  Europace       Date:  2021-05-25       Impact factor: 5.214

7.  Bleeding complications during cardiac electronic device implantation in patients receiving antithrombotic therapy: is there any value of local tranexamic acid?

Authors:  Osman Beton; Ersin Saricam; Hakki Kaya; Hasan Yucel; Orhan Dogdu; Okan Onur Turgut; Ocal Berkan; Izzet Tandogan; Mehmet Birhan Yilmaz
Journal:  BMC Cardiovasc Disord       Date:  2016-04-22       Impact factor: 2.298

8.  Biomarker-based diagnosis of pacemaker and implantable cardioverter defibrillator pocket infections: A prospective, multicentre, case-control evaluation.

Authors:  Carsten Lennerz; Hrvoje Vrazic; Bernhard Haller; Siegmund Braun; Tobias Petzold; Ilka Ott; Agnes Lennerz; Jonathan Michel; Patrick Blažek; Isabel Deisenhofer; Peter Whittaker; Christof Kolb
Journal:  PLoS One       Date:  2017-03-06       Impact factor: 3.240

9.  Effect of intraoperative local application of 3% hydrogen peroxide on pocket infections following cardiac implantable electronic device implantation: An observational study.

Authors:  Daniel Apel; Michele Brunelli; Mohammed El-Gahry; J Christoph Geller; Bernward Lauer; Marc-Alexander Ohlow
Journal:  Indian Pacing Electrophysiol J       Date:  2018-07-20
  9 in total