Is manual reticulocyte count a reliable option for under resourced countries?

Afsheen Farzand Ali1, Bushra Moiz, Sadia Omer.   

Abstract

OBJECTIVES: To establish the credibility of manual reticulocyte counts without compromising the quality of care, and to evaluate the degree of acceptability of manual reticulocyte counts in terms of accuracy and cost effectiveness in comparison with two automated haematology analyzers.
METHODS: Visual reticulocyte enumeration was evaluated for comparability, within-batch precision and costing with respect to Coulter STKS and Gen S haematology analyzers.
RESULTS: The results of reticulocyte estimation for 80 samples as obtained by 3 modes were correlated using Pearson's correlation coefficient (r) which were computed as 0.884, 0.875, and 0.793 for manual-Gen.S, Gen.SSTKS and manual-STKS respectively thus showing positive association of these results. STKS had the CV of 10.4% and was more precise compared to Coulter Gen.S (CV = 11.6%) while manual counts showed the least precision with a CV of 19.8%. The cost per test was calculated to be $0.11 for manual technique in contrast to $0.45 for Gen S and $1.09 for STKS.
CONCLUSION: Visual counting of reticulocytes can be used as a reliable tool for estimating reticulocytes in resource strained countries as it is not only cost effective but can also efficiently discriminate between high and low reticulocyte ranges which are required for sound clinical judgment.

Mesh:

Year:  2010        PMID: 21375188

Source DB:  PubMed          Journal:  J Pak Med Assoc        ISSN: 0030-9982            Impact factor:   0.781


  3 in total

1.  Improved light microscopy counting method for accurately counting Plasmodium parasitemia and reticulocytemia.

Authors:  Caeul Lim; Ligia Pereira; Pritish Shardul; Anjali Mascarenhas; Jennifer Maki; Jordan Rixon; Kathryn Shaw-Saliba; John White; Maria Silveira; Edwin Gomes; Laura Chery; Pradipsinh K Rathod; Manoj T Duraisingh
Journal:  Am J Hematol       Date:  2016-05-24       Impact factor: 10.047

2.  Effect of Lead Exposure on the Status of Reticulocyte Count Indices among Workers from Lead Battery Manufacturing Plant.

Authors:  Ravibabu Kalahasthi; Tapu Barman
Journal:  Toxicol Res       Date:  2016-10-30

3.  Giemsa-stained wet mount based method for reticulocyte quantification: a viable alternative in resource limited or malaria endemic settings.

Authors:  Wenn-Chyau Lee; Bruce Russell; Yee-Ling Lau; Mun-Yik Fong; Cindy Chu; Kanlaya Sriprawat; Rossarin Suwanarusk; Francois Nosten; Laurent Renia
Journal:  PLoS One       Date:  2013-04-02       Impact factor: 3.240

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.