Ho-Joong Jung1, Gautum Vangipuram, Matthew B Fisher, Guoguang Yang, Shanling Hsu, John Bianchi, Chad Ronholdt, Savio L-Y Woo.
Abstract
Soft tissue allografts, such as the bone-patellar tendon-bone (BPTB) graft, have been frequently used for anterior cruciate ligament (ACL) reconstruction. As allografts are subjected to freezing and thawing for multiple cycles, the objective of this study was to measure the changes of the biomechanical properties of the human BPTB allograft after 4 and 8 freeze-thaw cycles in comparison to a single freeze-thaw cycle. Three BPTB specimens were procured from 21 human donors and divided into three groups: 1, 4, or 8 freeze-thaw cycles. Each freeze-thaw cycle consisted of freezing at -20 ± 10°C for more than 6 h and thawing at 22 ± 3°C for at least 6 h. Tensile testing of the BPTB specimens consisted of loading between 50 N and 250 N for 100 cycles and then loading to failure. Cyclic loading revealed a similar amount of creep (∼0.5 mm) among the three freeze-thaw cycles groups (p = 0.38). The stiffness of the BPTB graft for the 1, 4, and 8 freeze-thaw cycle groups were 244 ± 42 N/mm, 235 ± 39 N/mm, and 231 ± 40 N/mm, respectively (p = 0.43). Similar findings were obtained for the ultimate load of the BPTB graft (p = 0.14) and the tangent modulus of the PT substance (p = 0.41). The results of this study suggest that there would be little measurable effect on the structural properties of the BPTB graft or mechanical properties of the PT tissue substance following 8 freeze-thaw cycles. These BPTB allografts could potentially be re-frozen without a loss in their biomechanical properties, given appropriate storage and care.Entities:
Mesh:
Year: 2011 PMID: 21374710 PMCID: PMC3132590 DOI: 10.1002/jor.21373
Source DB: PubMed Journal: J Orthop Res ISSN: 0736-0266 Impact factor: 3.494
Figure 1Typical BPTB specimen following isolation.
Figure 2(A) Experimental apparatus for tensile testing. The patellar and tibial bone blocks were gripped using custom clamps. Markers were placed on PT substance for strain tracking. (B) Typical load-elongation behavior of a BPTB specimen.
Gross Measurements of the PT Substance for the 1, 4, and 8 Freeze–Thaw Cycle Groups (Mean ± SD)
| Number of freeze–thaw cycles | |||
|---|---|---|---|
| 1 | 4 | 8 | |
| Length between bone blocks (mm) | 39.2 ± 7.7 | 37.8 ± 6.4 | 40.1 ± 7.0 |
| Width (mm) | 10.2 ± 0.7 | 10.3 ± 0.4 | 10.1 ± 0.3 |
| Thickness (mm) | 3.4 ± 0.7 | 3.4 ± 0.6 | 3.5 ± 0.5 |
| Cross-sectional area (mm2) | 49.9 ± 8.5 | 53.5 ± 7.5 | 52.1 ± 6.5 |
Effects of the Number of Freeze–Thaw Cycles on the Structural Properties of the BPTB Allografts (Mean ± SD)
| Number of freeze–thaw cycles | |||
|---|---|---|---|
| 1 ( | 4 ( | 8 ( | |
| (A) Parameters describing the structural properties of the BPTB | |||
| Creep (mm) | 0.4 ± 0.2 | 0.5 ± 0.3 | 0.6 ± 0.5 |
| Stiffness (N/mm) | 244 ± 42 | 235 ± 39 | 231 ± 40 |
| Ultimate load (N) | 1276 ± 322 | 1141 ± 319 | 1170 ± 357 |
| (B) Mode of failure (number of cases) | |||
| Bony avulsion | 9 (5 T, 4 P) | 6 (4 T, 2 P) | 5 (5 T, 0 P) |
| Detachment at tissue insertion to bone | 8 (8 T, 0 P) | 9 (9 T, 0 P) | 12 (10 T, 2 P) |
| Substance tear | 4 | 6 | 4 |
T, tibial side; P, patellar side.
Figure 3Graphical representation of the stiffness (A) and ultimate load (B) representing the structural properties of the BPTB grafts for the freeze–thaw cycle groups (mean ± SD).
Figure 4Graphical representation of the tangent modulus of the PT for the freeze–thaw cycle groups (mean ± SD).