Evaluation of expanded polytetrafluoroethylene-covered stents for the treatment of venous outflow stenosis in hemodialysis access grafts.

Michael G Chan1, Frank J Miller, Karim Valji, Michael D Kuo.   

Abstract

PURPOSE: To investigate the use of an expanded polytetrafluoroethylene-coated nitinol stent-graft in patients with failing hemodialysis access circuits and compare grafts with and without a heparin-coated surface.
MATERIALS AND METHODS: A single-institution retrospective review of 16 patients (average age, 55 years; seven men) treated for hemodialysis access circuit stenoses was performed. Patients had a prosthetic arteriovenous (AV) graft (n = 10) or autogenous AV fistula (n = 6), all with clinically relevant access stenoses that were refractory to angioplasty alone; 10 patients also had access circuit thrombosis at the time of treatment. Stents were placed in the basilic (n = 4), brachial (n = 7), cephalic (n = 4), and femoral (n = 1) veins. Study endpoints included primary treatment area patency, primary circuit patency, primary assisted circuit patency, and secondary circuit patency, followed to a maximum of 6 months.
RESULTS: Primary treatment area patency rates at 30 days and 180 days were 75% and 31%, respectively. Secondary patency rates at the same time points were 88% and 68%, respectively. Mean primary treatment area patency was 93 days. Thrombosis was a major cause of failure, occurring in seven of 11 cases of primary treatment area patency loss. Subgroup analysis demonstrated that the heparin-coated configuration (n = 7) showed a 6-month primary patency rate of 57%, compared with 11% for the stent-grafts without heparin bonding (n = 9; P = .06).
CONCLUSIONS: Overall, the majority of deployed stent-grafts lost primary patency within 6 months, with thrombosis a major cause of failure. However, the addition of a heparin-coated surface appeared to improve results in these otherwise treatment-refractory access circuits.
Copyright © 2011 SIR. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21371910     DOI: 10.1016/j.jvir.2010.12.013

Source DB:  PubMed          Journal:  J Vasc Interv Radiol        ISSN: 1051-0443            Impact factor:   3.464


  1 in total

1.  First Clinical Results of the Merit WRAPSODY™ Cell-Impermeable Endoprosthesis for Treatment of Access Circuit Stenosis in Haemodialysis Patients.

Authors:  James Gilbert; Jason Rai; David Kingsmore; John Skousen; Nikolaos Ptohis
Journal:  Cardiovasc Intervent Radiol       Date:  2021-09-12       Impact factor: 2.740

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.