Effect of air-flow on the evaluation of refractive surgery ablation patterns.

Carlos Dorronsoro1, Silvia Schumacher, Pablo Pérez-Merino, Jan Siegel, Michael Mrochen, Susana Marcos.   

Abstract

An Allegretto Eye-Q laser platform (Wavelight GmbH, Erlangen, Germany) was used to study the effect of air-flow speed on the ablation of artificial polymer corneas used for testing refractive surgery patterns. Flat samples of two materials (PMMA and Filofocon A) were ablated at four different air flow conditions. The shape and profile of the ablated surfaces were measured with a precise non-contact optical surface profilometer. Significant asymmetries in the measured profiles were found when the ablation was performed with the clinical air aspiration system, and also without air flow. Increasing air-flow produced deeper ablations, improved symmetry, and increased the repeatability of the ablation pattern. Shielding of the laser pulse by the plume of smoke during the ablation of plastic samples reduced the central ablation depth by more than 40% with no-air flow, 30% with clinical air aspiration, and 5% with 1.15 m/s air flow. A simple model based on non-inertial dragging of the particles by air flow predicts no central shielding with 2.3 m/s air flow, and accurately predicts (within 2 μm) the decrease of central ablation depth by shielding. The shielding effects for PMMA and Filofocon A were similar despite the differences in the ablation properties of the materials and the different full-shielding transmission coefficient, which is related to the number of particles ejected and their associated optical behavior. Air flow is a key factor in the evaluation of ablation patterns in refractive surgery using plastic models, as significant shielding effects are found with typical air-flow levels used under clinical conditions. Shielding effects can be avoided by tuning the air flow to the laser repetition rate.

Entities:  

Mesh:

Year:  2011        PMID: 21369297     DOI: 10.1364/OE.19.004653

Source DB:  PubMed          Journal:  Opt Express        ISSN: 1094-4087            Impact factor:   3.894


  2 in total

1.  Analytical optimization of the ablation efficiency at normal and non-normal incidence for generic super Gaussian beam profiles.

Authors:  Samuel Arba-Mosquera; Shwetabh Verma
Journal:  Biomed Opt Express       Date:  2013-07-24       Impact factor: 3.732

2.  Shielding effect of the smoke plume by the ablation of excimer lasers.

Authors:  Csaba Szekrényesi; Huba Kiss; Tamás Filkorn; Zoltán Zsolt Nagy
Journal:  BMC Ophthalmol       Date:  2018-10-23       Impact factor: 2.209

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.