Feasibility of using cone-beam CT to verify and reposition the optically guided target localization of linear accelerator based stereotactic radiosurgery.

Jingeng Zhu1.   

Abstract

PURPOSE: The optically guided target localization had been developed for linear accelerator based stereotactic radiosurgery (SRS). Unlike the traditional laser localization, the optical guided target localization utilizes a digital system to position patient. Although the system has been proven accurate and robust, it takes away the capability of physicist to directly double check the target position prior to irradiation. Any error from system calibration, data transformation, or head ring position maintenance will not be caught. The purpose of this work is to investigate the possibility of using cone-beam CT (CBCT) to double check the optically guided SRS target localization and reposition the patient.
METHODS: A SRS quality assurance (QA) phantom was used in the study. The phantom mounted with SRS head frame was scanned by computer tomography (CT) and planned according to the SRS radiation treatment planning process. A target isocenter is defined and transferred to the optically guided target localization system. The phantom was then transported to the linear accelerator room and localized at the initial position agreed by the optically guided target localization system and the CBCT system. Tests were conducted by moving/rotating the phantom to a set of preset offsets and taking CBCT images. Shifts detected by CBCT were compared with the preset offsets. Agreements between them were studied to see how well the CBCT was in discovering the optically guided target localization error.
RESULTS: Experiment results demonstrated good agreement between the CBCT detected phantom shift and the preset offset, when the offset is above 1 mm shift or 0.2 degree rotation. Offset less than 1 mm shift or 0.2 degree rotation was not detectable by CBCT.
CONCLUSIONS: The study concludes that the CBCT is able to discover the optically guided target localization error due to the system calibration or had ring migration. It is a valuable second check tool for SRS target localization quality assurance. The accuracy of CBCT in estimating patient positioning deviation satisfies the SRS procedures with generous tumor size and margin that can tolerate 1 mm or 0.2 degree accuracy. This avoids sending patient home without treatment. CBCT can be neither used as a primary SRS target localization nor can it be used to reposition the patient that cannot tolerate 1 mm shift or 0.2 degree rotation.

Entities:  

Mesh:

Year:  2011        PMID: 21361207     DOI: 10.1118/1.3531547

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  3 in total

1.  Optimization of an on-board imaging system for extremely rapid radiation therapy.

Authors:  Erica M Cherry Kemmerling; Meng Wu; He Yang; Peter G Maxim; Billy W Loo; Rebecca Fahrig
Journal:  Med Phys       Date:  2015-11       Impact factor: 4.071

2.  Prevention of gross setup errors in radiotherapy with an efficient automatic patient safety system.

Authors:  Guanghua Yan; Kathryn Mittauer; Yin Huang; Bo Lu; Chihray Liu; Jonathan G Li
Journal:  J Appl Clin Med Phys       Date:  2013-11-04       Impact factor: 2.102

3.  A protocol to extend the longitudinal coverage of on-board cone-beam CT.

Authors:  Dandan Zheng; Jun Lu; Ariel Jefferson; Cheng Zhang; Jian Wu; William Sleeman; Elisabeth Weiss; Nesrin Dogan; Shiyu Song; Jeffrey Williamson
Journal:  J Appl Clin Med Phys       Date:  2012-07-05       Impact factor: 2.102

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.